期刊论文详细信息
Sensors
Developing Barbed Microtip-Based Electrode Arrays for Biopotential Measurement
Li-Sheng Hsu1  Shu-Wei Tung1  Che-Hsi Kuo1 
[1] Department of Mechanical Engineering, National Taiwan University, Taipei 10617, Taiwan;
关键词: barbed microtips;    contact impedance;    detaching force;    dry electrode;    electrocardiography (ECG);    electroencephalogram (EEG);    silicon wet etching;   
DOI  :  10.3390/s140712370
来源: mdpi
PDF
【 摘 要 】

This study involved fabricating barbed microtip-based electrode arrays by using silicon wet etching. KOH anisotropic wet etching was employed to form a standard pyramidal microtip array and HF/HNO3 isotropic etching was used to fabricate barbs on these microtips. To improve the electrical conductance between the tip array on the front side of the wafer and the electrical contact on the back side, a through-silicon via was created during the wet etching process. The experimental results show that the forces required to detach the barbed microtip arrays from human skin, a polydimethylsiloxane (PDMS) polymer, and a polyvinylchloride (PVC) film were larger compared with those required to detach microtip arrays that lacked barbs. The impedances of the skin-electrode interface were measured and the performance levels of the proposed dry electrode were characterized. Electrode prototypes that employed the proposed tip arrays were implemented. Electroencephalogram (EEG) and electrocardiography (ECG) recordings using these electrode prototypes were also demonstrated.

【 授权许可】

CC BY   
© 2014 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190024052ZK.pdf 2158KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:14次