Sensors
Tracheal Sounds Acquisition Using Smartphones
Bersain A. Reyes1  Natasa Reljin1 
[1] Department of Biomedical Engineering, Worcester Polytechnic Institute, 100 Institute Road, Worcester, MA 01609, USA;
关键词: respiratory sounds;    tracheal sounds;    smartphone;    respiratory rate;    breath-phase;    entropy;    time-frequency representation;   
DOI  :  10.3390/s140813830
来源: mdpi
PDF
【 摘 要 】

Tracheal sounds have received a lot of attention for estimating ventilation parameters in a non-invasive way. The aim of this work was to examine the feasibility of extracting accurate airflow, and automating the detection of breath-phase onset and respiratory rates all directly from tracheal sounds acquired from an acoustic microphone connected to a smartphone. We employed the Samsung Galaxy S4 and iPhone 4s smartphones to acquire tracheal sounds from N = 9 healthy volunteers at airflows ranging from 0.5 to 2.5 L/s. We found that the amplitude of the smartphone-acquired sounds was highly correlated with the airflow from a spirometer, and similar to previously-published studies, we found that the increasing tracheal sounds' amplitude as flow increases follows a power law relationship. Acquired tracheal sounds were used for breath-phase onset detection and their onsets differed by only 52 ± 51 ms (mean ± SD) for Galaxy S4, and 51 ± 48 ms for iPhone 4s, when compared to those detected from the reference signal via the spirometer. Moreover, it was found that accurate respiratory rates (RR) can be obtained from tracheal sounds. The correlation index, bias and limits of agreement were r2 = 0.9693, 0.11 (−1.41 to 1.63) breaths-per-minute (bpm) for Galaxy S4, and r2 = 0.9672, 0.097 (–1.38 to 1.57) bpm for iPhone 4s, when compared to RR estimated from spirometry. Both smartphone devices performed similarly, as no statistically-significant differences were found.

【 授权许可】

CC BY   
© 2014 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190023493ZK.pdf 2802KB PDF download
  文献评价指标  
  下载次数:12次 浏览次数:9次