Sensors | |
Flexible Capacitive Electrodes for Minimizing Motion Artifacts in Ambulatory Electrocardiograms | |
Jeong Su Lee2  Jeong Heo2  Won Kyu Lee2  Yong Gyu Lim3  Youn Ho Kim1  | |
[1] Samsung Advanced Institute of Technology, 130, Samsung-ro, Yeongtong-gu, Suwon 443-803, Korea; E-Mail:;Interdisciplinary Program for Bioengineering, Graduate School, Seoul National University, Seoul 110-799, Korea; E-Mails:;Department of Oriental Biomedical Engineering, Sangji University, Wonju 220-702, Korea; E-Mail: | |
关键词: capacitive electrode; flexible electrode; ambulatory electrocardiogram; | |
DOI : 10.3390/s140814732 | |
来源: mdpi | |
【 摘 要 】
This study proposes the use of flexible capacitive electrodes for reducing motion artifacts in a wearable electrocardiogram (ECG) device. The capacitive electrodes have conductive foam on their surface, a shield, an optimal input bias resistor, and guarding feedback. The electrodes are integrated in a chest belt, and the acquired signals are transmitted wirelessly for ambulatory heart rate monitoring. We experimentally validated the electrode performance with subjects standing and walking on a treadmill at speeds of up to 7 km/h. The results confirmed the highly accurate heart rate detection capacity of the developed system and its feasibility for daily-life ECG monitoring.