期刊论文详细信息
Sensors
Rapid Detection of Microorganisms Based on Active and Passive Modes of QCM
Zdeněk Farka1  David Kovář1 
[1]CEITEC MU, Masaryk University, Kamenice 5, 625 00 Brno, Czech Republic
[2] E-Mails:
关键词: Escherichia coli;    immunosensor;    impedance analysis;    label-free detection;    quartz crystal microbalance;   
DOI  :  10.3390/s150100079
来源: mdpi
PDF
【 摘 要 】

Label-free immunosensors are well suited for detection of microorganisms because of their fast response and reasonable sensitivity comparable to infection doses of common pathogens. Active (lever oscillator and frequency counter) and passive (impedance analyzer) modes of quartz crystal microbalance (QCM) were used and compared for rapid detection of three strains of E. coli. Different approaches for antibody immobilization were compared, the immobilization of reduced antibody using Sulfo‐SMCC was most effective achieving the limit of detection (LOD) 8 × 104 CFU·mL−1 in 10 min. For the passive mode, software evaluating impedance characteristics in real-time was developed and used. Almost the same results were achieved using both active and passive modes confirming that the sensor properties are not limited by the frequency evaluation method but mainly by affinity of the antibody. Furthermore, reference measurements were done using surface plasmon resonance. Effect of condition of cells on signal was observed showing that cells ruptured by ultrasonication provided slightly higher signal changes than intact microbes.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190018221ZK.pdf 1470KB PDF download
  文献评价指标  
  下载次数:13次 浏览次数:20次