期刊论文详细信息
Sensors
Enhanced Visualization of Fine Needles Under Sonographic Guidance Using a MEMS Actuator
Zhiyuan Shen1  Yufeng Zhou1  Jianmin Miao1 
[1] School of Mechanical and Aerospace Engineering, Nanyang Technological University, 50 Nanyang Ave., Singapore 639798, Singapore; E-Mails:
关键词: biopsy;    ultrasound imaging;    acoustic interference;    MEMS actuator;   
DOI  :  10.3390/s150203107
来源: mdpi
PDF
【 摘 要 】

Localization of a needle tip is important for biopsy examinations in clinics. However, the needle tip is sometimes difficult to discern under the guidance of sonography due to its poor visibility. A mini actuator that radiates a low-intensity ultrasound wave was manufactured using micro-electro-mechanical system (MEMS) technology. Interference between the radiated and diagnostic ultrasound pulses was observed as bright lines in the B-mode ultrasound image, from which the mini actuator could be recognized with ease. Because the distance between the mini actuator and the needle tip is fixed, the needle tip can be determined despite its inconsistent appearance in the sonography. Both gel phantom and ex vivo tissue evaluation showed that the needle tip can be determined reliably utilizing the acoustic interference pattern.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190016797ZK.pdf 1247KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:27次