期刊论文详细信息
Sensors
An Improved Electrochemical Aptasensor for Chloramphenicol Detection Based on Aptamer Incorporated Gelatine
Ezat Hamidi-Asl2  Freddy Dardenne1  Ronny Blust1  Karolien De Wael2 
[1] Sphere Research Group, Department of Biology, University of Antwerp, Groenenborgerlaan 171, B-2020 Antwerp, Belgium; E-Mails:;AXES Research Group, Department of Chemistry, University of Antwerp, Groenenborgerlaan 171, B-2020 Antwerp, Belgium; E-Mail:
关键词: aptamers;    gelatine;    chloramphenicol;    electrochemical sensor;   
DOI  :  10.3390/s150407605
来源: mdpi
PDF
【 摘 要 】

Because of the biocompatible properties of gelatine and the good affinity of aptamers for their targets, the combination of aptamer and gelatine type B is reported as promising for the development of biosensing devices. Here, an aptamer for chloramphenicol (CAP) is mixed with different types of gelatine and dropped on the surface of disposable gold screen printed electrodes. The signal of the CAP reduction is investigated using differential pulse voltammetry. The diagnostic performance of the sensor is described and a detection limit of 1.83 × 10−10 M is found. The selectivity and the stability of the aptasensor are studied and compared to those of other CAP sensors described in literature.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190014770ZK.pdf 3408KB PDF download
  文献评价指标  
  下载次数:16次 浏览次数:16次