期刊论文详细信息
Sensors
Performance Evaluation of Smartphone Inertial Sensors Measurement for Range of Motion
Quentin Mourcou1  Anthony Fleury2  Céline Franco1  Frຝéric Klopcic2  Nicolas Vuillerme1 
[1] University of Grenoble-Alpes, AGIM, La Tronche 38700, France; E-Mails:;University of Lille, F-59000 Lille and Mines Douai, URIA, Douai F-59508, France; E-Mails:
关键词: smartphone sensing;    IMU;    Kalman filter;    validation;   
DOI  :  10.3390/s150923168
来源: mdpi
PDF
【 摘 要 】

Over the years, smartphones have become tools for scientific and clinical research. They can, for instance, be used to assess range of motion and joint angle measurement. In this paper, our aim was to determine if smartphones are reliable and accurate enough for clinical motion research. This work proposes an evaluation of different smartphone sensors performance and different manufacturer algorithm performances with the comparison to the gold standard, an industrial robotic arm with an actual standard use inertial motion unit in clinical measurement, an Xsens product. Both dynamic and static protocols were used to perform these comparisons. Root Mean Square (RMS) mean values results for static protocol are under 0.3° for the different smartphones. RMS mean values results for dynamic protocol are more prone to bias induced by Euler angle representation. Statistical results prove that there are no filter effect on results for both protocols and no hardware effect. Smartphones performance can be compared to the Xsens gold standard for clinical research.

【 授权许可】

CC BY   
© 2015 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190006304ZK.pdf 459KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:6次