期刊论文详细信息
Acoustical Science and Technology
Uncertainty analysis for pressure sensitivities of laboratory standard microphones
Ryuzo Horiuchi1  Sojun Sato1  Takeshi Fujimori1 
[1] National Metrology Institute of Japan, AIST
关键词: Laboratory standard microphone;    Pressure sensitivity;    Pressure calibration;    Reciprocity technique;    Measurement uncertainty;   
DOI  :  10.1250/ast.25.354
学科分类:声学和超声波
来源: Acoustical Society of Japan
PDF
【 摘 要 】

References(10)Cited-By(4)Pressure sensitivities of laboratory standard microphones are acoustic standards required for precise measurement in the audible frequency range. The National Metrology Institute of Japan (NMIJ) has developed a pressure calibration system that enabled absolute calibration of pressure sensitivities. In this study, we analyzed the measurement uncertainty of pressure sensitivities calibrated at NMIJ in detail by using the pressure sensitivity formula and examining every possible component that contributed to the uncertainty. The analysis revealed several dominant components, such as repeatability in the measurement of the voltage transfer function, cavity volume of the coupler, capillary tube correction factor, and instability of the microphone. Within a frequency range in which the expanded uncertainty was constant, the expanded uncertainty (coverage factor k=2) was 0.04 dB for Brüel & Kjær 4160, a type LS1P microphone, and 0.09 dB for Brüel & Kjær 4180, a type LS2aP microphone.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080715440ZK.pdf 211KB PDF download
  文献评价指标  
  下载次数:12次 浏览次数:23次