期刊论文详细信息
Metrology and Measurement Systems
Using A Segmented Voltage Sweep Mode and A Gaussian Curve Fitting Method to Improve Heavy Metal Measurement System Performance
Octavian Postolache1  Pedro Girão1  José Pereira1 
关键词: heavy metals measurement;    electrochemical analysis;    voltammetry;    auto-calibration and curve fitting;   
DOI  :  10.2478/v10178-012-0033-8
来源: Versita
PDF
【 摘 要 】

This paper presents a voltammetric segmented voltage sweep mode that can be used to identify and measure heavy metals' concentrations. The proposed sweep mode covers a set of voltage ranges that are centered around the redox potentials of the metals that are under analysis. The heavy metal measurement system can take advantage of the historical database of measurements to identify the metals with higher concentrations in a given geographical area, and perform a segmented sweep around predefined voltage ranges or, alternatively, the system can perform a fast linear voltage sweep to identify the voltammetric current peaks and then perform a segmented voltage sweep around the set of voltages that are associated with the voltammetric current peaks. The paper also includes the presentation of two auto-calibration modes that can be used to improve system's reliability and proposes the usage of a Gaussian curve fitting of voltammetric data to identify heavy metals and to evaluate their concentrations. Several simulation and experimental results, that validate the theoretical expectations, are also presented in the paper.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080706885ZK.pdf 794KB PDF download
  文献评价指标  