期刊论文详细信息
JOURNAL OF CHEMICAL ENGINEERING OF JAPAN
Measurement of Vapor–Liquid Equilibria for Binary Systems Using a Hold-Up Compensable Ebulliometer
Kiyofumi Kurihara1  Shigemichi Uno1  Hiroyuki Matsuda1  Kenji Ochi1  Shinichiro Kikkawa1  Katsumi Tochigi1 
[1] Department of Materials and Applied Chemistry, College of Science and Technology, Nihon University
关键词: Vapor Hold-Up;    Vapor–Liquid Equilibria;    Ebulliometer;    Experimental;    Correlation;   
DOI  :  10.1252/jcej.40.288
来源: Maruzen Company Ltd
PDF
【 摘 要 】

References(32)Cited-By(5)This study presents a new ebulliometer which is able to compensate for the vapor hold-up problem commonly experienced when measuring vapor–liquid equilibria. Firstly, the influence of the vapor hold-up was examined by measuring the boiling points for the ethanol + toluene system at 53.33 kPa, and it was confirmed that the vaporization coefficient of the newly-developed ebulliometer was 2.2%. Secondly, the boiling points of three binary systems of ethanol + toluene, ethanol + cyclohexane, and benzene + cyclohexane, were determined in the range of 53.33 to 98.66 kPa. In addition, the experimental boiling point data were satisfactorily correlated using the NRTL equation. Finally, the determined NRTL parameters were evaluated by calculating the vapor–liquid equilibria under the same conditions as the thermodynamically reliable values in the literature.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080696056ZK.pdf 19KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:9次