| Química Nova | |
| Evaluation of the corrosion resistance of stainless steel ISO 5832-9 and F138 implants removed from patients | |
| Oliveira, Luiz Fernando Cappa de1  Silva, Elison da Fonseca e1  Instituto Federal de Educação Ciência e Tecnologia Sudeste de Minas Gerais, Juiz de Fora, Brasil1  Universidade Federal de Juiz de Fora, Juiz de Fora, Brasil1  | |
| 关键词: stain steel; orthopedic implant; corrosion.; | |
| DOI : 10.1590/S0100-40422012000100012 | |
| 学科分类:化学(综合) | |
| 来源: Sociedade Brasileira de Quimica | |
PDF
|
|
【 摘 要 】
This work has compared the surfaces of two different steel samples used as orthopedical implants, classified as ASTM F138 and ISO5832-9, through optical emission spectroscopy, by means of SEM and EDS. The samples (implants) were also submitted to potentiodynamic cyclic polarization in Ringer lactate and NaCl 0.9 M L-1 solutions; ISO5832-9 sample did not show any kind of localized corrosion, but in the case of F138 steel was observed a pit localized corrosion in both solutions. In Ringer lactate solution it was observed a loss of about 63% for nickel and 26% for iron for F138 stell, compared to the initial composition.