This work presents a new approach to control the flow rate in hydrodynamic flow experiments. The combination of air pressure generated by an aquarium air pump and the pressure generated by a water column were used for this purpose. This device supports a stable flow rate without pulsation for a long period of time. Furthermore, the flow rate can be easily controlled at various values in one or more streams. The performance of this approach was investigated using Fe(CN)64- solutions in flowing systems using amperometric and voltammetric detection in wall-jet configuration. The results showed that the performance of the proposed device was better than a commercial peristaltic pump. It suggests that this approach can be used successfully in flow analysis systems.