期刊论文详细信息
Química Nova
Atomic and molecular threshold photoelectron spectroscopy
Silva, Helen1  Universidade Federal de Juiz de Fora, Juiz de Fora1  Lopes, Maria Cristina Andreolli1  Moreira, Cleber Dias1  Couto, Helder1 
关键词: threshold photoelectron spectroscopy;    synchrotron radiation;    photoionization.;   
DOI  :  10.1590/S0100-40422006000100028
学科分类:化学(综合)
来源: Sociedade Brasileira de Quimica
PDF
【 摘 要 】

A threshold photoelectron spectrometer applied to the study of atomic and molecular threshold photoionization processes is described. The spectrometer has been used in conjunction with a toroidal grating monochromator at the National Synchrotron Radiation Laboratory (LNLS), Brazil. It can be tuned to accept threshold electrons (< 20 meV) and work with a power resolution of 716 (~18 meV at 12 eV) with a high signal/noise ratio. The performance of this apparatus and some characteristics of the TGM (Toroidal Grating Monochromator) beam line of LNLS are described and discussed by means of argon, O2 and N2 threshold photoelectron spectra.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050593466ZK.pdf 287KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:9次