期刊论文详细信息
Journal of the Brazilian Chemical Society
Sequential voltammetric determination of chromium, thallium, cadmium, lead, copper and antimony in saline hemodialysis concentrates using electrolyte pH gradient
Nascimento, Paulo C1  Jost, Cristiane L1  Bohrer, Denise1  Carvalho, Leandro M. de1  Schneider, Alexandre B1  Universidade Federal de Santa Maria, Santa Maria, Brazil1 
关键词: sequential voltammetry;    dialysate concentrates;    trace metals;    contamination;   
DOI  :  10.1590/S0103-50532011000500003
学科分类:化学(综合)
来源: SciELO
PDF
【 摘 要 】

The sequential voltammetric determination of chromium, thallium, cadmium, lead, copper and antimony in dialysate concentrates (DCs), is presented herein. Chromium was quantified by adsorptive cathodic stripping voltammetry and thallium, cadmium, lead, copper and antimony were assayed by anodic stripping voltammetry. The novel method developed to test these samples exploits the effect of electrolyte pH gradient on voltammetric response. The limits of detection (LOD) ranged from 0.03 µg L-1 for cadmium to 0.27 µg L-1 for copper. Recoveries from spiked samples were 92.0-117.5%. This method was applied to the analysis of commercial DCs, where the investigated metals were found in some samples at concentrations between 0.16 µg L-1 for antimony and 140.00 µg L-1 for copper. Interferences and the reference method are discussed.