Journal of the Brazilian Chemical Society | |
Development of a fluorimetric sequential injection analysis (SIA) methodology for determination of quinine | |
Infante, Carlos M. C.1  Masini, Jorge C.1  Universidade de São Paulo,, São Paulo, Brazil1  | |
关键词: sequential injection analysis; quinine; fluorimetry; dispersion; optimization; | |
DOI : 10.1590/S0103-50532011001000009 | |
学科分类:化学(综合) | |
来源: SciELO | |
【 摘 要 】
A sequential injection analysis (SIA) methodology was developed for fluorimetric determination of quinine in soft drinks. Minimum dispersion and efficient mixing were achieved by aspirating 200 mL of sample between two 100 mL reagent zones (0.10 mol L-1 H2SO4) at flow rate of 250 mL s-1 and using a reaction coil length of 50 cm (0.8 mm internal diameter). Linear response for quinine concentrations (C) between 0.050 and 100.0 mg L-1 was described by: I = (532 ± 40) + (2.36 ± 0.04) C, with r = 0.999, where I is the relative fluorescence intensity. The coefficient of variation and limits of detection and quantification were 1.9% (0.50 mg L-1 n = 10), 2.3 and 4.5 mg L-1, respectively. The sampling throughput was 60 analyses per hour, consuming 2.2 mL of concentrated H2SO4 and producing 4 mL of wastes per analysis. No statistically significant differences were observed between the results obtained by the proposed SIA methodology and the ones obtained by the batch procedure.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201912050581159ZK.pdf | 306KB | download |