期刊论文详细信息
Journal of the Brazilian Chemical Society
Determination of gibberellic acid by sequential injection analysis using a potentiometric detector based on Mn(III)-porphyrin with improved characteristics
Couto, Cristina M. C. M.1  Faculty of Pharmacy, Porto, Portugal1  Santos, Emília M. G.1  Montenegro, M. Conceição B. S. M.1  Universidade de São Paulo, Piracicaba, Brazil1  ISCS, Paredes, Portugal1  Araújo, Alberto N.1  Reis, Boaventura F.1 
关键词: gibberellic acid;    ion selective tubular electrode;    SIA;    metalloporphyrins;    agricultural growth promoters;   
DOI  :  10.1590/S0103-50532004000500015
学科分类:化学(综合)
来源: SciELO
PDF
【 摘 要 】

This work describes the determination of gibberellic acid in growth promoters formulations used in agriculture by sequential injection analysis and potentiometric detection. The potentiometric detectors with improved characteristics used a PVC membrane prepared with Mn(III)tetraphenylporphyrin-Cl as electroactive specie. Different membranes formulated with several ionic additives were compared in order to select the most suitable one concerning slope, response time, reproducibility and selectivity. A membrane containing 6 wt% of manganese tetraphenylporphyrin-Cl [(Mn(TPP)Cl], 27 wt% of PVC, 66 wt% of dibutylphthalate (DBP), as mediator solvent, and 1 wt% of sodium tetraphenylborate, as anionic additive, was chosen for analytical applications. When coupled to a SIA system the tubular electrode presented a linear range between 5×10-4 mol L-1 and 8×10-3 mol L-1, a slope of -64.5 ± 1.6 mV/dec and lower potentiometric selectivity coefficients (log Kpot) than those obtained with the conventional electrodes. Standard deviation of 0.01 and 0.4 (n= 4) and a sampling rate of 30 samples/hour were obtained in the analysis of real samples.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050579591ZK.pdf 206KB PDF download
  文献评价指标  
  下载次数:13次 浏览次数:9次