期刊论文详细信息
Journal of the Brazilian Chemical Society
Determination of fluoride in coal using pyrohydrolysis for analyte separation
Universidade Federal do Rio Grande do Sul, Porto Alegre, Brazil1  Flores, Érico M. M.1  Paniz, José N. G.1  Flores, Éder L. M.1  Pozebon, Dirce1  Dressler, Valderi L.1  Universidade Federal de Santa Maria, Santa Maria, Brazil1 
关键词: coal;    fluoride;    pyrohydrolysis;    ion electrode;    vanadium pentoxide;    catalyst;   
DOI  :  10.1590/S0103-50532003000200024
学科分类:化学(综合)
来源: SciELO
PDF
【 摘 要 】

A method for the determination of fluoride in coal using pyrohydrolysis for sample decomposition is proposed. An apparatus was constructed, and the influence of vanadium pentoxide (V2O5) as a catalyst for the pyrohydrolysis reaction was investigated. It was verified that V2O5 does not influence the release of fluoride from coal. Certified and non-certified samples were analysed using an ion selective electrode (ISE) for analyte detection. Precise (RSD < 7%) and accurate (recovery in accordance with certified values) results were obtained. The limit of quantification (LOQ) of the method was 5.0 mgF/g of sample, using 20 mg of sample and a final dilution to 10 mL. The sample frequency was 5 samples h-1.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050579390ZK.pdf 88KB PDF download