Journal of the Brazilian Chemical Society | |
Analytical performance of an inductively coupled plasma optical emission spectrometry with Dual View configuration | |
Universidade Estadual de Campinas, Campinas, Brazil1  Universidade Federal de São Carlos, São Carlos, Brazil1  Baccan, Nivaldo1  Silva, Júlio C. J.1  Nóbrega, Joaquim A.1  | |
关键词: ICP OES; axial view; radial view; robustness; sensitivity; Mg II/Mg I ratio; | |
DOI : 10.1590/S0103-50532003000200020 | |
学科分类:化学(综合) | |
来源: SciELO | |
【 摘 要 】
The analytical performance of axially and radially viewed inductively coupled plasma optical emission spectrometers (AX-ICP OES and RD-ICP OES, respectively) were evaluated in terms of the Mg II/Mg I ratio, matrix effects, and detection limits (LOD). All solutions were made in either 1% v v-1 nitric acid or in a 10% v v-1 water-soluble tertiary amine solution (CFA-C) adjusted to pH 8. Using a nebulization gas flow-rate of 0.5 L min-1 and applied power of 1.3 kW, Mg II/Mg I ratio was at least 15 in any medium for both configurations. Under robust conditions, plasma processes were practically not affected by the presence of 0.1% m v-1 of Ca, K, or Na in both configurations. However, interference effects were observed with non-robust conditions, mainly with AX-ICP OES in both media. The best detection limits were obtained using AX-ICP OES under robust conditions, with solutions prepared in the amine medium.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201912050579386ZK.pdf | 149KB | download |