期刊论文详细信息
Journal of Orthopaedics and Traumatology
Posterior dislocation of the Oxford knee meniscal bearing: a treatment option
P. McLardy-Smith2  S. B. Tibrewal3  D. W. Murray2  H. Pandit2  S. Tibrewal1 
[1] St Bartholomew’s and the Royal London Hospitals$$;Nuffield Orthopaedic Centre$$;Queen Elizabeth Hospital$$
关键词: Oxford knee;   
DOI  :  10.1007/s10195-013-0250-2
学科分类:骨科学
来源: Springer
PDF
【 摘 要 】

Unicompartmental knee replacement (UKR) is now established as a treatment for medial compartment arthritis. The Oxford UKR (Biomet Orthopedics, Inc, Warsaw, IN, USA) has a mobile-bearing system, which minimizes wear. This has been shown to provide excellent long-term results. Dislocation of the mobile-bearing device is rare with an incidence of 1 in 200 (0.5 %). The treatment usually involves exploration of the knee through the original anteromedial incision, removal of the dislocated bearing and rectification of the underlying cause for the dislocation. We describe two cases of a posterior dislocation in which the mobile bearing could not be retrieved and was left in situ. In both cases a good outcome was achieved. We conclude that in extremely rare cases where a dislocated bearing has migrated posteromedially and cannot be retrieved, it can be left in place rather than exploring the joint acutely through a separate posterior incision.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040500017ZK.pdf 315KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:20次