Neutron detector array at IUAC: Design features and instrumentation developments | |
J Antony1  I M Govil1  Rajesh Kumar1  N Saneesh1  T Varughese1  J Zacharias1  K S Golda1  P Barua1  A Jhingan1  K Singh1  R K Bhowmik1  V V Satyanarayana1  A Kothari1  S K Saini1  R P Singh1  S K Suman1  S Venkataramanan1  S K Mandal1  P Sugathan11  Ruby Shanti1  A Gupta1  B R Behera1  | |
[1] Inter University Accelerator Centre, Aruna Asaf Ali Marg, New Delhi 110 067, India$$ | |
关键词: Fusion–fission reactions; instrumentation; scintillation detectors.; | |
DOI : | |
学科分类:物理(综合) | |
来源: Indian Academy of Sciences | |
【 摘 要 】
The characteristics and performance of the newly commissioned neutron detector array at IUAC are described. The array consists of 100 BC501 liquid scintillators mounted in a semispherical geometry and are kept at a distance of 175 cm from the reaction point. Each detector is a 5″ × 5″ cylindrical cell coupled to 5″ diameter photomultiplier tube (PMT). Signal processing is realized using custom-designed home-made integrated electronic modules which perform neutron–gamma discrimination using zero cross timing and time-of-flight (TOF) technique. Compact custom-built high voltage power supply developed using DC–DC converters are used to bias the detector. The neutrons are recorded in coincidence with fission fragments which are detected using multi-wire proportional counters mounted inside a 1m diameter SS target chamber. The detectors and electronics have been tested off-line using radioactive sources and the results are presented.