期刊论文详细信息
Pramana
Performance studies on high pressure 1-D position sensitive neutron detectors
S S Desai1  A M Shaikh11 
[1] Solid State Physics Division, Bhabha Atomic Research Centre, Mumbai 400 085, India$$
关键词: Gas-filled proportional counters;    position sensitive detectors;    neutron scattering and diffraction.;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

The powder diffractometer and Hi-Q diffractometer at Dhruva reactor make use of five identical 1-D position sensitive detectors (PSDs) to scan scattering angles in the range 3° to 140°. In order to improve the overall throughput of these spectrometers, it is planned to install a bank of 15 high-efficiency and high-resolution PSDs arranged in three layers with five PSDs in each layer. With each high pressure PSD (3He 10 bar + Kr 2 bar) showing the efficiency gain of 1.8 at 1.2 Å, detector bank is expected to show overall gain of 5.5 times the present detection efficiency and reduction in data collection time by equivalent factor. The 1-D PSDs are developed in batches of five, and are characterized so that all PSDs operate at uniform parameters such as position resolution, uniformity of efficiency and linearity of response. Position spectrum indicates the differential position resolution to be ∼ 1 mm and integral position resolution to be 3–4 mm. Broadening of position spectrum at the extreme end of sensitive length of PSD is analysed using fine shift of the beam. Dependence of position resolution and dynamic range of output pulse on the input impedance of pre-amplifier is also presented.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040497762ZK.pdf 767KB PDF download
  文献评价指标  
  下载次数:16次 浏览次数:23次