期刊论文详细信息
Pramana
Instrumentation for PSD-based neutron diffractometers at Dhruva reactor
S P Borkar1  P K Mukhopadhyay1  V D Srivastava1  S K Kataria1  M D Ghodgaonkar1  S S Pande11  S Prafulla1  A Behare1 
[1] Electronics Division, Bhabha Atomic Research Centre, Mumbai 400 085, India$$
关键词: Data acquisition system;    position sensitive detector;    ratio ADC.;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

Linear position sensitive detectors (PSDs) are widely used to configure neutron diffractometers and other instruments. Necessary front-end electronics and a data acquisition system [1] is developed to cater to such instruments built around the Dhruva research reactor in BARC. These include three diffractometers with multiple PSDs and four with single PSD. The front-end electronics consists of high voltage units, preamplifiers [2], shaping amplifiers, ratio ADCs (RDC) [3]. The data acquisition system consists of an interface card and software. Commercially available hardware like temperature controller or stepper motor controller connected over GPIB or RS232 are also integrated in the data acquisition system. The data acquisition is automated so that it can continue unattended for control parameter like temperature, thus enabling optimum utilization of available beam time. The instrumentation is scalable and can be easily configured for various instrumental requirements. The front-end electronics and the data acquisition system are described here.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040496690ZK.pdf 568KB PDF download