A possibility of parallel and anti-parallel diffraction measurements on neutron diffractometer employing bent perfect crystal monochromator at the monochromatic focusing condition
Pivel Mikula21  Sung Baek Kim2  Sung Kyu Kim2  Yong Nam Choi2  Shin Ae Kim2  Chang-Hee Lee2 
[1] Nuclear Physics Institute, 25068 Rez, Czech Republic$$;HANARO Center, Korea Atomic Energy Research Institute, Yuseong, Daejeon 305-600, Korea$$
关键词: Neutron diffractometer;    bent crystal monochromator;    monochromatic focusing.;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

In a conventional diffractometer having single monochromator, only one position, parallel position, is used for the diffraction experiment (i.e. detection) because the resolution property of the other one, anti-parallel position, is very poor. However, a bent perfect crystal (BPC) monochromator at monochromatic focusing condition can provide a quite flat and equal resolution property at both parallel and anti-parallel positions and thus one can have a chance to use both sides for the diffraction experiment. From the data of the FWHM and the 𝛥 𝑑/𝑑 measured on three diffraction geometries (symmetric, asymmetric compression and asymmetric expansion), we can conclude that the simultaneous diffraction measurement in both parallel and anti-parallel positions can be achieved.