期刊论文详细信息
Pramana
Tokamak Plasmas : Internal magnetic field measurement in tokamak plasmas using a Zeeman polarimeter
M Jagadeeshwari1  J Govindarajan1 
[1] Institute for Plasma Research, Bhat Gandhinagar 382 428, India$$
关键词: Tokamak;    polarimeter;    Zeeman effect;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

In a tokamak plasma, the poloidal magnetic field profile closely depends on the current density profile. We can deduce the internal magnetic field from the analysis of circular polarization of the spectral lines emitted by the plasma. The theory of the measurement and a detailed design of the Zeeman polarimeter constructed to measure the poloidal field profile in the ADITYA tokamak are presented. The Fabry-Perot which we have employed in our design, with photodiode arrays followed by lock-in detection of the polarization signal, allows the measurement of the fractional circular polarization. In this system He-II line with wavelength 4686 Å is adopted as the monitoring spectral line. The line emission used in the present measurement is not well localized in the plasma, necessiating the use of a spatial inversion procedure to obtain the local values of the field.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040495721ZK.pdf 78KB PDF download
  文献评价指标  
  下载次数:19次 浏览次数:42次