Bulletin of the Korean chemical society | |
A Study on the Peak Separation of Acetone and Acrolein Based on High-Performance Liquid Chromatography (HPLC) Method | |
Chang Hwan Kim1  Jin Su Park1  Shin Do Kim1  Jeong Joo Lee1  | |
关键词: Acetone; Acrolein; Tetrahydrofuran; Peak separation; HPLC; | |
DOI : | |
学科分类:化学(综合) | |
来源: Korean Chemical Society | |
【 摘 要 】
To resolve the differentiation problem of acetone and acrolein in the analysis of carbonyls by high-performance liquid chromatography (HPLC), we investigated the optimum analytical conditions for their separation. Carbonyl compounds were collected by 2,4-dinitrophenylhydrazine (DNPH)-coated cartridges. We examined the influence of three experimental variables: temperature (25, 30, 40, 50 and 60 oC), flow rate (1.0 and 1.2 mL/min), and relative mobile phase composition (among acetonitrile, water and tetrahydrofuran). The experimental results revealed the optimum analytical condition of a flow rate of 1.2 mL/min, temperature of 32 oC and mobile phase composition of acetonitrile: water: tetrahydrofuran = 34 : 52.8 : 13.2. The analysis of indoor air composition indicated that acrolein and acetone comprised 11% and 42% of all aldehydes, respectively.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201912010241759ZK.pdf | 385KB | download |