| Bulletin of the Korean chemical society | |
| Electrochemical Assay of Neurotransmitter Glycine in Brain Cells | |
| Suw Young Ly1  Dong Young Kim1  | |
| 关键词: Voltammetry; Neurotransmitter; Glycine; Fish; Brain; | |
| DOI : | |
| 学科分类:化学(综合) | |
| 来源: Korean Chemical Society | |
PDF
|
|
【 摘 要 】
Neurotransmitter glycine in the nano gram range was analyzed using a paste electrode (PE) in cyclic voltammetry (CV) and square-wave stripping voltammetry (SWSV). An anodic peak caused by oxidation of the glycine ion appeared at the 0.4 V (versus Ag/AgCl/KCl) potential in a 0.1 M NH4H2PO4 electrolyte solution. At optimized conditions, the working range of the SWSV and CV concentration was found to be 5-60 ngL-1 glycine; precision of R2 = 0.9816 (SWSV) and 0.9986 (CV); and detection limit of 0.65 ngL-1 (5.82 �? 10-12 molL-1) (S/N = 3). The optimized conditions were applied to an assay in a fish brain tissue and a living brain cell in real time.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912010240461ZK.pdf | 311KB |
PDF