| Bulletin of the Korean chemical society | |
| Analytical Head-space Supercritical Fluid Extraction Methodology for the Determination of Organochlorine Compounds in Aqueous Matrix | |
| Won Kyoung Lee1  Keon Sang Ryoo1  Yong Pyo Hong1  Yonggyun Kim1  Seong-Oon Ko1  Jong-Ha Choi1  | |
| 关键词: Head-space SFE methodology; Organochlorine compounds; Conventional method; Static and dynamic mode; | |
| DOI : | |
| 学科分类:化学(综合) | |
| 来源: Korean Chemical Society | |
PDF
|
|
【 摘 要 】
The proposed head-space supercritical fluid extraction (SFE) methodology as an alternative to an existing conventional procedure was explored for the determination of organochlorine compounds in aqueous matrix. In this study, polychlorinated biphenyls (PCBs) and organochlorine pesticides (OCPs) were utilized as target analytes. To enhance the recovery efficiency, the factors such as the CO2 density, the extraction time, and the extraction mode were investigated. Furthermore, the analytical procedures and the results obtained were compared with those provided by the conventional method (the U.S. EPA method 8080). Under the optimized conditions, i.e., a combination of static with dynamic SFE mode at 2,000 psi and 40 oC, the head-space SFE methodology gave equivalent or better to the conventional method in recovery efficiencies with clear advantages such as simple sample treatment and fast analysis time as well as reduced solvent and reagent consumption.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912010240086ZK.pdf | 2352KB |
PDF