期刊论文详细信息
Journal of Nuclear Medicine
Evaluation of the Genisys4, a Bench-Top Preclinical PET Scanner
Johannes Czernin1  Ken Herrmann1  Arion Chatziioannou1  Liu Wei1  Caius Radu1  David Nathanson1  Magnus Dahlbom1 
关键词: preclinical PET;    18F-FDG;    tumor models;    scanner performance;   
DOI  :  10.2967/jnumed.112.114926
学科分类:医学(综合)
来源: Society of Nuclear Medicine
PDF
【 摘 要 】

The Genisys4 is a small bench-top preclinical PET scanner designed to enable imaging in biology, biochemistry, and pharmacology laboratories and imaging centers. Here, we compare its performance with that of a well-established preclinical PET scanner. Methods: Subcutaneous and lung tumor xenografts were used to compare lesion detectability and treatment responses to chemotherapy (gemcitabine) using 18F-FDG PET. The size of subcutaneous xenografts (L1210 and L1210-10K leukemia cells) and lung metastases (B-16 melanoma cells) was measured on small-animal CT images. Tumor 18F-FDG uptake was expressed as percentage injected dose per gram. Using list-mode data, serial images of the left ventricular blood pool were used to generate time–activity curves. Results: Subcutaneous xenografts (range, 4–12 mm; mean ± SD, 6.1 ± 1.7 mm) and lung metastases (range, 1–5 mm; mean, 2.1 ± 1.2 mm) were detected equally well with both scanners. Tumor 18F-FDG uptake measured with both scanners was highly correlated for subcutaneous xenografts (r2 = 0.93) and lung metastases (r2 = 0.83). The new Genisys4 scanner and the established scanner provided comparable treatment response information (r2 = 0.93). Dynamic imaging sequences permitted the generation of left ventricular blood-pool time–activity curves with both scanners. Conclusion: Using subcutaneous and lung xenografts, a novel and an established preclinical PET scanner provided equivalent information with regard to lesion detection, tumor 18F-FDG uptake, tumor response to treatment, and generation of time–activity curves. Thus, the Genisys4 provides a small, efficient bench-top preclinical PET alternative for quantitatively studying murine tumor models in biology, biochemistry, and pharmacology laboratories and preclinical imaging centers.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010198809ZK.pdf 861KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:29次