| Photomultiplier Tube Artifacts on 67Ga-Citrate Imaging Caused by Loss of Correction Floods Due to an Off-Peak Status of One Head of a Dual-Head γ-Camera | |
| Na Song1  Ruth Lorenzo1  Charito Love1  Sridivya Jaini1  Joseph E. Glaser1  | |
| 关键词: gamma camera; scintigraphy; quality control; electronic off-peak; artifact; gallium; | |
| DOI : 10.2967/jnmt.111.097246 | |
| 学科分类:医学(综合) | |
| 来源: Society of Nuclear Medicine | |
PDF
|
|
【 摘 要 】
γ-cameras use flood-field corrections to ensure image uniformity during clinical imaging. A loss or corruption of the correction data of one head of a dual-head camera can result in an off-peak artifactual appearance. We present our experience with the occurrence of such an incident on a 67Ga scan. Methods: A patient was referred for a whole-body 67Ga scan to evaluate for causes of neutropenic fever. Whole-body planar and static images of the head, chest, abdomen, pelvis, and lower extremities in multiple projections were obtained. Results: Whole-body images showed decreased image quality on the anterior view obtained with detector 1 and an unremarkable posterior image obtained with detector 2. A problem with detector 2 was suspected, and additional static images were obtained after rotation of the detector heads. The posterior images taken with detector 1 showed photomultiplier tube outlines. The anterior images taken with detector 2 showed improved count and image quality. It was later found that the uniformity map for detector 2 had been lost and that this software malfunction led to the resulting imaging problem. Conclusion: When artifacts with an off-peak appearance are seen on scintigraphic images, evaluation of possible causes should include not only isotope window settings but also an incorrect or corrupted uniformity map.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912010179790ZK.pdf | 247KB |
PDF