期刊论文详细信息
Journal of Nuclear Medicine Technology
Review of the Linogram and Sinogram: An Easy Way to Detect Off-Peak Artifacts in Myocardial Perfusion SPECT
Seyed Rasoul Zakavi1  Vahid Reza Dabbagh Kakhki1  Mehdi Mommennezhad1  Ramin Sadeghi1 
[1] Department of Nuclear Medicine, Emam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran Department of Nuclear Medicine, Emam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran Department of Nuclear Medicine, Emam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran
关键词: sinogram;    artifact;    myocardial perfusion SPECT;    photopeak;   
DOI  :  10.2967/jnmt.108.061218
学科分类:医学(综合)
来源: Society of Nuclear Medicine
PDF
【 摘 要 】

The correct energy window setting and a stable photopeak position are crucial for good data acquisition. We report a patient whose myocardial perfusion SPECT images showed mildly irregular radiotracer uptake when the patient was prone but not when the patient was supine. Review of the sinogram showed clear interruptions in the sinusoidal shape of the raw data when the patient was prone. Detailed technical evaluation revealed that the artifact was caused by an unstable photopeak position due to malfunction of a power supply board of the γ-camera. This case emphasizes the importance of reviewing the sinogram display and raw data to assess the integrity of data acquisition and to reveal any potential technical problems such as a spontaneous shift of the photopeak.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010179670ZK.pdf 499KB PDF download
  文献评价指标  
  下载次数:7次 浏览次数:1次