Journal of Nuclear Medicine Technology
Comparison of Methods of Attenuation and Scatter Correction in Brain Perfusion SPECT
Masayasu Takeuchi1  Jun Deguchi1  Isamu Narabayashi1  Keita Utsunomiya1  Masuo Hayashi1  Makoto Yamada1  Tsuyoshi Komori1  Kensei Kanna1 
关键词: brain perfusion SPECT;    autoradiography method;    stable XeCT regional CBF measurement;    scatter correction;    attenuation correction;   
DOI  :  
学科分类:医学(综合)
来源: Society of Nuclear Medicine
PDF
【 摘 要 】

Objective: The radioactivity count distribution in the brain must be determined accurately to accurately measure cerebral blood flow (CBF). Scatter and attenuation are factors that compromise the accuracy of determining radioactivity counts in the brain.Methods: We compared regional CBF in patients by an autoradiographic method using N-isopropyl-p-[123I] iodoamphetamine when, first, attenuation correction alone was performed uniformly on SPECT images by using empiric μ-values (Chang method); second, scatter correction was performed and the μ-values of a homogeneous-attenuation body of water were used for attenuation correction (triple-energy window [TEW] + Chang method); and third, scatter correction was performed and the μ-values calculated by CT were used for attenuation correction (TEW + CT method). We also compared regional CBF measured by these methods with the values obtained by the xenon CT/CBF method, which uses CT and stable xenon.Results: Scatter correction reduced overestimation of regional CBF in low-flow regions. The TEW + CT method yielded better regional and overall correlations with the xenon CT/CBF method than did either of the other methods.Conclusion: The TEW + CT method of correction gave the most accurate measurements of regional CBF.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912010179561ZK.pdf 523KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:3次