A PC- Based Transient Method for Thermal Conductivity Measurement.
A.K. Singh1 
[1] Institute of Armament Technology, Pune
关键词: Propellant oxidisers;    Thermal conductivity;    Automated measurements;    Thermal probes;   
DOI  :  
学科分类:社会科学、人文和艺术(综合)
来源: Defence Scientific Information & Documentation Centre
PDF
【 摘 要 】

In this paper, an indigenously developed thermal probe has been interfaced with a PC for automated measurement of thermal conductivity (K ). The developed system has been calibrated and standardised by measuring K of glycerol. The maximum percentage error, for repeated sets of observations, was within 7.29 per cent of standard value reported for glycerol. This methodology has been successfully employed for measuring K of propellant oxidisers, additives, binders, etc.