期刊论文详细信息
IEICE Electronics Express
A method for reducing body exposure to electromagnetic field of pillow type wireless charger in fully implantable middle ear hearing device
J. H. Lee2  E. S. Jung1  M. W. Kim1  J. H. Cho1  K. W. Seong1  J. W. Lee1  H. G. Lim1 
[1] Graduate School of Electrical Engineering and Computer Science, Kyungpook National University;Department of Biomedical Engineering, Kyungpook National University Hospital
关键词: fully implantable middle ear hearing device;    wireless charger;    electromagnetic coupling;    electromagnetic field exposure;    hall effect sensor;   
DOI  :  10.1587/elex.6.1318
学科分类:电子、光学、磁材料
来源: Denshi Jouhou Tsuushin Gakkai
PDF
【 摘 要 】

References(10)Implantable middle ear hearing devices (IMEHDs) have been developed at several countries. Some devices can be selected rechargeable battery for long-term implant. In South Korea, a kind of IMEHD with rechargeable battery and a pillow type wireless charger have been developed. The charger includes several coils surrounding user's head in sleep. However, this device should excessively consume power due to driving all primary coils. Especially, needless exposure to magnetic field generated from uncoupled primary coils may induce bio-hazards. Therefore, a method reducing the body exposure to electromagnetic field was proposed as decreasing the emitted total magnetic flux during charge process.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201911300348741ZK.pdf 1953KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:16次