Analysis of the effect of stent porosity and shape on saccular intracranial aneurysm using the Lattice Boltzmann method
Ayala, D. S1  Madero, D. A2 
[1] Oncólogos del Occidente, Armenia, Colombia;Clínica de Occidente S.A., Armenia, Colombia
关键词: saccular intracranial aneurysm;    stent;    lattice boltzmann method;   
DOI  :  
学科分类:工程和技术(综合)
来源: Universidad E A F I T
PDF
【 摘 要 】

This article presents an analysis of blood flow patterns in intracranial saccular aneurysm and the effects of the shape and porosity of the stents used in endovascular treatments. In this study will be used the flow reduction criteria for characterizing the efficiency of the stent. The hemodynamic properties of a newtonian blood flow into the aneurysm will be evaluated using the Lattice Boltzmann method (LBM). Porosity values and stent forms are proposed for analysis. In all stent cases analyzed is observed a reduction of velocity and pressure and an increase in viscosity. It is further noted that the rectangular contour stent is the optimal case and reduces the magnitude of the flow velocity inside the aneurysm much as 76%. The results help to understand the role of porosity in the form and design of a stent.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201911300316680ZK.pdf 3206KB PDF download