期刊论文详细信息
Oftal mokhirurgiya
ПРОЧНОСТНЫЕ ХАРАКТЕРИСТИКИ РОГОВИЦЫ КРОЛИКАПОСЛЕ ИНТРАСТРОМАЛЬНОЙ ИМПЛАНТАЦИИ РАЗЛИЧНЫХ ИМПЛАНТОВС ПРИМЕНЕНИЕМ ФЕМТОСЕКУНДНОГО ЛАЗЕРА
Н. А. Поздеева1  М. В. Синицын2  И. Н. Лакомкина3  Н. П. Паштаев4 
[1] ГАУ ЧР ДПО «Институт усовершенствования врачей» Минздрава Чувашской Республики;ФГБОУ ВО Чувашский государственный университет имени И.Н. Ульянова;ГАУ ЧР ДПО «Институт усовершенствования врачей» Минздрава Чувашской Республики;ФГАУ «НМИЦ «МНТК «Микрохирургия глаза» им. акад. С.Н. Федорова» Минздрава России, Чебоксарский филиал
关键词: intra-stromal closing ring with a «lock»;    MyoRing;    intrastromal segment;    the femtosecond laser;    интрастромальное смыкающееся кольцо с «замком»;    кольцо MyoRing;    интрастромальный сегмент;    фемтосекундный лазер;   
DOI  :  10.25276/0235-4160-2018-2-42-45
学科分类:眼科学
来源: M H T K Mikrokhirurgiya Glaza
PDF
【 摘 要 】

Purpose.   Comparative  experimental  study  of  the  effect  of  the implantation of various intrastromal implants using a femtosecond laser (FSL)on the strength properties of the rabbit cornea. Material and methods . The experimental work was performed on 24 corneas of isolated eyes of 12 rabbits enucleated 1 month after the slaughter. All eyes were divided into 4 groups, depending on the method of operation. Formation of intra-stromal pockets (IP) and tunnels (IT) was performed using the FSL Femto Visum. The eyes of rabbits with intact corneas were included into the first control group. In the group 2, the intrastromal segment 350° was implanted into the IT. In the group 3,the MyoRing was implanted into the IP. In the group 4, the intrastromal closing ring with a «lock» was implanted into the IT. Results.After the implantation of intrastromal implants in comparison with the control group it was noted that the strength of the examined cornea of rabbits was increased to break. The greatest increase in the strength properties of the rabbit cornea was observed after the implantation of the intrastromal closing ring with a «lock» in comparison with the implantation of the intrastromal segment of 350° and the MyoRing. Conclusions.Implantation of intrastromal rings and segments leads  to  an  increase  in  the  strength  properties  of  the  cornea.Implantation of an intrastromal closing ring with a «lock» leads to the greatest increase in the strength properties of the cornea compared with the implantation of the intrastromal segment of 350° and the MyoRing.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO201910253241946ZK.pdf 193KB PDF download