期刊论文详细信息
Pramana: Journal of physics
Analytical evaluation of cesium emission lines using laser-induced breakdown spectroscopy
MANJEET SINGH^1,21  ARNAB SARKAR^1,22 
[1] Fuel Chemistry Division, Bhabha Atomic Research Centre, Mumbai 400 094, India^1;Homi Bhabha National Institute, Anushaktinagar, Mumbai 400 094, India^2
关键词: Laser-induced breakdown spectroscopy;    emission;    spectroscopy;    cesium;    linear dynamic range;   
DOI  :  
学科分类:物理(综合)
来源: Indian Academy of Sciences
PDF
【 摘 要 】

A laser-induced breakdown spectroscopy-based method has been successfully developed to quantify cesium (Cs) in solution using spectroscopically pure graphite planchets as a sample support. As Cs is a line-poor system, only five usable Cs atomic emission lines could be found and characterised by employing high-resolution system. The calibration curves of these emission lines were constructed under optimised experimental conditions. The analytical properties of these calibration curves were evaluated based on the usable dynamic range, $R^{2}$ of fitting, root mean square error cross-validation and limit of detection (LOD). The dynamic ranges of these five lines were found to be in correlation with the energy level involved in the transition. An LOD of 4 ppm was obtained using Cs(I) 852.11-nm line, which corresponds to 0.16 $\mu$g of Cs on the planchet. Based on the cross-validation approach, the best accuracy and precision ($\sim$6%) were obtained for 852.11 nm in >3000 ppm solutions, and the same is $\sim$8% for 672.33 nm and 697.33 nm in high concentrated solution of Cs.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO201910251278538ZK.pdf 190KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:9次