iPhone 4s Photoplethysmography: Which Light Color Yields the Most Accurate Heart Rate and Normalized Pulse Volume Using the iPhysioMeter Application in the Presence of Motion Artifact?
Kenta Matsumura1  Peter Rolfe2  Takehiro Yamakoshi3  Jihyoung Lee4 
[1] College of Science and Engineering, Kanazawa University, Kanazawa, Ishikawa, Japan;Department of Automatic Measurement and Control, Harbin Institute of Technology, Harbin, Heilongjiang, China;Graduate School of Natural Science and Technology, Kanazawa University, Kanazawa, Ishikawa, Japan;Oxford BioHorizons Ltd., Maidstone, Kent, United Kingdom
关键词: Cell phones;    Electrocardiography;    Motion;    Light pulses;    Heart rate;    Light;    Arterioles;    Cameras;   
DOI  :  10.1371/journal.pone.0091205
学科分类:医学(综合)
来源: Public Library of Science
PDF
【 摘 要 】

Recent progress in information and communication technologies has made it possible to measure heart rate (HR) and normalized pulse volume (NPV), which are important physiological indices, using only a smartphone. This has been achieved with reflection mode photoplethysmography (PPG), by using a smartphone’s embedded flash as a light source and the camera as a light sensor. Despite its widespread use, the method of PPG is susceptible to motion artifacts as physical displacements influence photon propagation phenomena and, thereby, the effective optical path length. Further, it is known that the wavelength of light used for PPG influences the photon penetration depth and we therefore hypothesized that influences of motion artifact could be wavelength-dependant. To test this hypothesis, we made measurements in 12 healthy volunteers of HR and NPV derived from reflection mode plethysmograms recorded simultaneously at three different spectral regions (red, green and blue) at the same physical location with a smartphone. We then assessed the accuracy of the HR and NPV measurements under the influence of motion artifacts. The analyses revealed that the accuracy of HR was acceptably high with all three wavelengths (all rs > 0.996, fixed biases: −0.12 to 0.10 beats per minute, proportional biases: r = −0.29 to 0.03), but that of NPV was the best with green light (r = 0.791, fixed biases: −0.01 arbitrary units, proportional bias: r = 0.11). Moreover, the signal-to-noise ratio obtained with green and blue light PPG was higher than that of red light PPG. These findings suggest that green is the most suitable color for measuring HR and NPV from the reflection mode photoplethysmogram under motion artifact conditions. We conclude that the use of green light PPG could be of particular benefit in ambulatory monitoring where motion artifacts are a significant issue.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO201904022678794ZK.pdf 1810KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:41次