| Proceedings of Singapore Healthcare | |
| Assessing Performance of i-STAT at the Point of Care in the Emergency Room: | |
| Chin PinYeo1  | |
| 关键词: competency; evaluation; near-patient testing; | |
| DOI : 10.1177/201010581102000304 | |
| 学科分类:医学(综合) | |
| 来源: Sage Journals | |
PDF
|
|
【 摘 要 】
Context:Point-of-care testing (POCT) for blood gas and chemistries is widely adopted in the emergency department (ED) for fast turnaround and critical medical decisions.Objective:To assess performance of ED physicians compared to laboratory technologists carrying out i-STAT analysis. Impact of user skill on POCT performance was also studied.Design:Over a 3-month period, ED physicians performed tests with their i-STAT unit with parallel blood specimens (n=60) sent to the central laboratory to be tested on another i-STAT unit and laboratory analysers (blood gas and chemistries). Comparisons to laboratory results (obtained with ABL 520 and Beckman-Coulter LX20PRO) were carried out. Two operators from ED and central laboratory performed side by side, repeated tests on the i-STAT units. Their overall within-run imprecision were compared. The analytical performance of i-STAT was further assessed by another laboratory technologist to verify its capability to deliver good results.Results:Platform comparisons show...