期刊论文详细信息
卷:215
Rapid microplate, green method for high-throughput evaluation of vinegar acidity using thermal infrared enthalpimetry
Tischer, Bruna ; Oliveira, Alessandra Stangherlin ; Ferreira, Daniele de Freitas ; Menezes, Cristiano Ragagnin ; Duarte, Fabio Andrei ; Wagner, Roger ; Barin, Juliano Smanioto
Univ Fed Santa Maria
关键词: Food;    Infrared thermography;    Imaging;    Calorimetry;    Enthalpimetric analysis;    Green analytical chemistry;   
DOI  :  10.1016/j.foodchem.2016.07.127
学科分类:食品科学和技术
PDF
【 摘 要 】
Infrared thermal imaging was combined with disposable microplates to perform enthalpimetric analysis using an infrared camera to monitor temperature without contact. The proposed thermal infrared enthalpimetry (TIE) method was used to determine the total, fixed and volatile acidities of vinegars. Sample preparation and analysis were performed in the same vessel, avoiding excessive sample handling and reducing energy expenditure by more than ten times. The results agreed with those of the conventional method for different kinds of vinegars, with values of 1.7%, and 23% for repeatability and intermediate precision, respectively. A linear calibration curve was obtained from 0.040 to 1.30 mol L-1. The proposed method provided rapid results (within 10 s) for four samples simultaneously, a sample throughput of up to 480 samples per hour. In addition, the method complies with at least eight of twelve recommendations for green analytical chemistry, making TIE a promising tool for routine vinegar analysis. (C) 2016 Elsevier Ltd. All rights reserved.
【 授权许可】

   

【 预 览 】
附件列表
Files Size Format View
JA201706070000860SK.pdf KB PDF download
  文献评价指标  
  下载次数:15次 浏览次数:54次