期刊论文详细信息
Proceedings
Development and Evaluation of Tri-Axial Fiber Bragg Grating Force Sensor for Catheter
Shin, Dongjoo1 
关键词: FBG;    force sensor;    optical fiber;    catheter;   
DOI  :  10.3390/proceedings1040549
学科分类:社会科学、人文和艺术(综合)
来源: mdpi
PDF
【 摘 要 】

Abstract: Avoiding the electromagnetic interference is one of the most difficult technical challenge during catheterization. The use of an optical fiber force sensor in the catheter can be an alternative. Herein, two flexure structures were designed for catheter tip and simulated using ANSYS to verify their sensitivity and durability. Fiber Bragg Grating sensors were configured to measure the tri-axial force exerted on the catheter tip. Calibrated result shows that an optical fiber force sensor can measure an axial force of up to 50 gF with an accuracy of 1 gF. Our calibration results demonstrate reliable measurement with high accuracy with repeatability.