Journal of Astronomical Instrumentation | |
MuSICa: THE MULTI-SLIT IMAGE SLICER FOR THE EST SPECTROGRAPH | |
K. KUEHN1  R. HUPE2  | |
[1] Argonne National Laboratory, Lemont, IL 60439, USA;Department of Physics, The Ohio State University, Columbus, OH 43210, USA | |
关键词: Integral field unit; image slicer; European Solar Telescope; integral field spectrograph; | |
DOI : 10.1142/S2251171713500098 | |
学科分类:天文学(综合) | |
来源: World Scientific Publishing Co. Pte. Ltd. | |
【 摘 要 】
Integral field spectroscopy (IFS) is a technique that allows one to obtain the spectra of all the points of a bidimensional field of view simultaneously. It is being applied to the new generation of the largest night-time telescopes but it is also an innovative technique for solar physics. This paper presents the design of a new image slicer, MuSICa (Multi-Slit Image slicer based on collimator-Camera), for the integral field spectrograph of the 4-m aperture European Solar Telescope (EST). MuSICa is a multi-slit image slicer that decomposes an 80 arcsec2 field of view into slices of 50 μm and reorganizes it into eight slits of 0.05 arcsec width × 200 arcsec length. It is a telecentric system with an optical quality at diffraction limit compatible with the two modes of operation of the spectrograph: spectroscopic and spectro-polarimetric. This paper shows the requirements, technical characteristics and layout of MuSICa, as well as other studied design options.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201902015339200ZK.pdf | 1289KB | download |