期刊论文详细信息
Analytical methods
Highly sensitive detection of copper(I) and copper(II) in cell specimens by CE-UV with large-volume sample stacking
Xin Li1  Zhiwei Zhu1  Meixian Li1  Xia Xie1  Fengxia Chang1 
[1] Beijing National Laboratory for Molecular Sciences, College of Chemistry and Molecular Engineering, Peking University,Beijing 100871,P. R. China
DOI  :  10.1039/C6AY00933F
学科分类:分析化学
来源: Royal Society of Chemistry
PDF
【 摘 要 】

A sensitive, rapid and simple CE-UV method with large-volume sample stacking (LVSS) and polarity switching has been developed and validated for the analysis of copper(I) and copper(II) simultaneously. Benefiting from the combination of the LVSS method, a limit of detection (LOD) of 30 nM was obtained which was much lower than those of the current detection methods. Several factors, such as the concentration and pH of the background solution, injection time and pressure, stacking voltage and inversion electric current, were systematically optimized. Under the optimum conditions, the detection limits of copper(I) and copper(II) were improved nearly 1000 times compared with the conventional methods by CE-UV, which made it possible to determine trace copper in cell specimens. Eventually, the proposed method was successfully applied to detect copper(I) and copper(II) in spiked cell samples with good recoveries...

【 授权许可】

CC BY-NC   

【 预 览 】
附件列表
Files Size Format View
RO201901236492538ZK.pdf 1047KB PDF download
  文献评价指标  
  下载次数:4次 浏览次数:0次