期刊论文详细信息
Journal of Rehabilitation and Assistive Technologies Engineering
Development of a step counting algorithm using the ambulatory tibia load analysis system for tibia fracture patients:
AradLajevardi-Khosh1 
关键词: Biomedical devices;    community ambulation;    continuous fracture care;    gait;    patient behaviour monitoring devices;    rehabilitation;    rehabilitation devices;    step activity monitor;    underfoot load monitoring;   
DOI  :  10.1177/2055668318804974
学科分类:工程和技术(综合)
来源: Sage Journals
PDF
【 摘 要 】

Introduction: Ambulation can be used to monitor the healing of lower extremity fractures. However, the ambulatory behavior of tibia fracture patients remains unknown due to an inability to continuously quantify ambulation outside of the clinic. The goal of this study was to design and validate an algorithm to assess ambulation in tibia fracture patients using the ambulatory tibial load analysis system during recovery, outside of the clinic.MethodsData were collected from a cyclic tester, 14 healthy volunteers performing a 2-min walk test on the treadmill, and 10 tibia fracture patients who wore the ambulatory tibial load analysis system during recovery.ResultsThe algorithm accurately detected 2000/2000 steps from simulated ambulatory data. During the 2-min walk test, step counts derived from the algorithm and treadmill showed a strong correlation (r2>0.98) to the visual (“actual”) step count. Applying the algorithm to continuous data from tibia fracture patients revealed qualitative differences in gait be...

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO201901224846852ZK.pdf 593KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:12次