| Journal of Rehabilitation and Assistive Technologies Engineering | |
| Development and reliability of a hand-held dynamometer device to measure the ankle dorsiflexor muscle strength: | |
| MasanobuKimura1  | |
| 关键词: Ankle dorsiflexor muscle strength; reliability testing; h; -held dynamometer; inter-rater; intra-rater; compensatory motion; | |
| DOI : 10.1177/2055668318802558 | |
| 学科分类:工程和技术(综合) | |
| 来源: Sage Journals | |
PDF
|
|
【 摘 要 】
BackgroundAnkle dorsiflexor muscle strength is a crucial component of gait.ObjectiveWe describe the development of a simple, hand-held dynamometer to measure the ankle dorsiflexor muscle strength in the sitting position. In addition, we examine its intra- and inter-rater reliability.MethodsMeasurements of the peak ankle dorsiflexor muscle strength were obtained by two examiners for 30 ankles of 15 healthy adults at two time points, with a one-day interval between measurements, to determine the inter- and intra-rater reliability. The intraclass correlation coefficients were calculated, and an intraclass correlation coefficient > 0.90 was considered as excellent reliability. A Bland–Altman analysis was used to assess systemic bias. The minimal detectable change in muscle strength was calculated with a confidence level of 95% (MDC95).ResultsThe reliability of the device was excellent for both intra- (intraclass correlation coefficients [1,3] = 0.94) and inter-rater (intraclass correlation coefficients [2,3] ...
【 授权许可】
CC BY
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201901222779583ZK.pdf | 209KB |
PDF