Radiation Oncology
Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors
Matthias Guckenberger1  Michael Flentje1  Gerd Müller1  Vanessa Homann1  Silke Stark1  Benedikt Seubert1  Reinhart A Sweeney1 
[1] Department of Radiation Oncology, University of Wuerzburg, Josef-Schneider-Str, 11 97080 Wuerzburg, Germany
关键词: Respiration correlated imaging;    Inter-observer variability;    Cone-beam CT;    Image-guidance;    Lung cancer;   
Others  :  1160835
DOI  :  10.1186/1748-717X-7-81
 received in 2012-04-16, accepted in 2012-06-08,  发布年份 2012
PDF
【 摘 要 】

Background

To analyze the accuracy and inter-observer variability of image-guidance (IG) using 3D or 4D cone-beam CT (CBCT) technology in stereotactic body radiotherapy (SBRT) for lung tumors.

Materials and methods

Twenty-one consecutive patients treated with image-guided SBRT for primary and secondary lung tumors were basis for this study. A respiration correlated 4D-CT and planning contours served as reference for all IG techniques. Three IG techniques were performed independently by three radiation oncologists (ROs) and three radiotherapy technicians (RTTs). Image-guidance using respiration correlated 4D-CBCT (IG-4D) with automatic registration of the planning 4D-CT and the verification 4D-CBCT was considered gold-standard. Results were compared with two IG techniques using 3D-CBCT: 1) manual registration of the planning internal target volume (ITV) contour and the motion blurred tumor in the 3D-CBCT (IG-ITV); 2) automatic registration of the planning reference CT image and the verification 3D-CBCT (IG-3D). Image quality of 3D-CBCT and 4D-CBCT images was scored on a scale of 1–3, with 1 being best and 3 being worst quality for visual verification of the IGRT results.

Results

Image quality was scored significantly worse for 3D-CBCT compared to 4D-CBCT: the worst score of 3 was given in 19 % and 7.1 % observations, respectively. Significant differences in target localization were observed between 4D-CBCT and 3D-CBCT based IG: compared to the reference of IG-4D, tumor positions differed by 1.9 mm ± 0.9 mm (3D vector) on average using IG-ITV and by 3.6 mm ± 3.2 mm using IG-3D; results of IG-ITV were significantly closer to the reference IG-4D compared to IG-3D. Differences between the 4D-CBCT and 3D-CBCT techniques increased significantly with larger motion amplitude of the tumor; analogously, differences increased with worse 3D-CBCT image quality scores. Inter-observer variability was largest in SI direction and was significantly larger in IG using 3D-CBCT compared to 4D-CBCT: 0.6 mm versus 1.5 mm (one standard deviation). Inter-observer variability was not different between the three ROs compared to the three RTTs.

Conclusions

Respiration correlated 4D-CBCT improves the accuracy of image-guidance by more precise target localization in the presence of breathing induced target motion and by reduced inter-observer variability.

【 授权许可】

   
2012 Sweeney et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150411081444530.pdf 1192KB PDF download
Figure 4 . 89KB Image download
Figure 3 . 52KB Image download
Figure 2 . 85KB Image download
Figure 1 . 66KB Image download
【 图 表 】

Figure 1 .

Figure 2 .

Figure 3 .

Figure 4 .

【 参考文献 】
  • [1]Seppenwoolde Y, Shirato H, Kitamura K, Shimizu S, van Herk M, Lebesque JV, Miyasaka K: Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. Int J Radiat Oncol Biol Phys 2002, 53(4):822-834.
  • [2]Hurkmans CW, van Lieshout M, Schuring D, van Heumen MJ, Cuijpers JP, Lagerwaard FJ, Widder J, van der Heide UA, Senan S: Quality Assurance of 4D-CT Scan Techniques in Multicenter Phase III Trial of Surgery Versus Stereotactic Radiotherapy (Radiosurgery Or Surgery for operable Early stage (Stage 1A) non-small-cell Lung cancer [ROSEL] Study). Int J Radiat Oncol Biol Phys 2010, 80(3):918-927.
  • [3]Richter A, Wilbert J, Baier K, Flentje M, Guckenberger M: Feasibility study for markerless tracking of lung tumors in stereotactic body radiotherapy. Int J Radiat Oncol Biol Phys 2010, 78(2):618-627.
  • [4]Guckenberger M, Krieger T, Richter A, Baier K, Wilbert J, Sweeney RA, Flentje M: Potential of image-guidance, gating and real-time tracking to improve accuracy in pulmonary stereotactic body radiotherapy. Radiother Oncol 2009, 91(3):288-295.
  • [5]Korreman S, Persson G, Nygaard D, Brink C, Juhler-Nottrup T: Respiration-correlated image guidance is the most important radiotherapy motion management strategy for most lung cancer patients. Int J Radiat Oncol Biol Phys 2012. EPUB ahead of print
  • [6]Sonke JJ, Zijp L, Remeijer P, van Herk M: Respiratory correlated cone beam CT. Med Phys 2005, 32(4):1176-1186.
  • [7]Hugo GD, Liang J, Campbell J, Yan D: On-line target position localization in the presence of respiration: a comparison of two methods. Int J Radiat Oncol Biol Phys 2007, 69(5):1634-1641.
  • [8]Guckenberger M, Meyer J, Wilbert J, Baier K, Mueller G, Wulf J, Flentje M: Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumors. Acta Oncol 2006, 45(7):897-906.
  • [9]Purdie TG, Bissonnette JP, Franks K, Bezjak A, Payne D, Sie F, Sharpe MB, Jaffray DA: Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys 2007, 68(1):243-252.
  • [10]Masi L, Casamassima F, Menichelli C, Pasciuti K, Doro R, Polli C, D'Imporzano E, Bonucci I: On-line image guidance for frameless stereotactic radiotherapy of lung malignancies by cone beam CT: comparison between target localization and alignment on bony anatomy. Acta Oncol 2008, 47(7):1422-1431.
  • [11]Wang Z, Wu QJ, Marks LB, Larrier N, Yin FF: Cone-beam CT localization of internal target volumes for stereotactic body radiotherapy of lung lesions. Int J Radiat Oncol Biol Phys 2007, 69(5):1618-1624.
  • [12]Bissonnette JP, Franks KN, Purdie TG, Moseley DJ, Sonke JJ, Jaffray DA, Dawson LA, Bezjak A: Quantifying interfraction and intrafraction tumor motion in lung stereotactic body radiotherapy using respiration-correlated cone beam computed tomography. Int J Radiat Oncol Biol Phys 2009, 75(3):688-695.
  • [13]Sonke JJ, Rossi M, Wolthaus J, van Herk M, Damen E, Belderbos J: Frameless stereotactic body radiotherapy for lung cancer using four-dimensional cone beam CT guidance. Int J Radiat Oncol Biol Phys 2009, 74(2):567-574.
  • [14]Grills IS, Hugo G, Kestin LL, Galerani AP, Chao KK, Wloch J, Yan D: Image-Guided Radiotherapy via Daily Online Cone-Beam CT Substantially Reduces Margin Requirements for Stereotactic Lung Radiotherapy. Int J Radiat Oncol Biol Phys 2008, 70(4):1045-1056.
  • [15]Worm ES, Hansen AT, Petersen JB, Muren LP, Praestegaard LH, Hoyer M: Inter- and intrafractional localisation errors in cone-beam CT guided stereotactic radiation therapy of tumours in the liver and lung. Acta Oncol 2010, 49(7):1177-1183.
  • [16]Guckenberger M, Meyer J, Wilbert J, Richter A, Baier K, Mueller G, Flentje M: Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors. Radiother Oncol 2007, 83(1):57-64.
  • [17]Potters L, Kavanagh B, Galvin JM, Hevezi JM, Janjan NA, Larson DA, Mehta MP, Ryu S, Steinberg M, Timmerman R, et al.: American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guideline for the performance of stereotactic body radiation therapy. Int J Radiat Oncol Biol Phys 2010, 76(2):326-332.
  • [18]Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, Keall P, Lovelock M, Meeks S, Papiez L, et al.: Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys 2010, 37(8):4078-4101.
  • [19]Guckenberger M, Sweeney RA, Wilbert J, Krieger T, Richter A, Baier K, Mueller G, Sauer O, Flentje M: Image-guided radiotherapy for liver cancer using respiratory-correlated computed tomography and cone-beam computed tomography. Int J Radiat Oncol Biol Phys 2008, 71(1):297-304.
  • [20]Shirato H, Suzuki K, Sharp GC, Fujita K, Onimaru R, Fujino M, Kato N, Osaka Y, Kinoshita R, Taguchi H, et al.: Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy. Int J Radiat Oncol Biol Phys 2006, 64(4):1229-1236.
  • [21]Korreman S, Rasch C, McNair H, Verellen D, Oelfke U, Maingon P, Mijnheer B, Khoo V: The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide. Radiother Oncol 2010, 94(2):129-144.
  • [22]Potters L, Gaspar LE, Kavanagh B, Galvin JM, Hartford AC, Hevezi JM, Kupelian PA, Mohiden N, Samuels MA, Timmerman R, et al.: American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guidelines for image-guided radiation therapy (IGRT). Int J Radiat Oncol Biol Phys 2010, 76(2):319-325.
  • [23]Galerani AP, Grills I, Hugo G, Kestin L, Mohammed N, Chao KK, Suen A, Martinez A, Yan D: Dosimetric impact of online correction via cone-beam CT-based image guidance for stereotactic lung radiotherapy. Int J Radiat Oncol Biol Phys 2010, 78(5):1571-1578.
  文献评价指标  
  下载次数:68次 浏览次数:15次