| Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency department | |
| Bernd Saugel2  Roland M Schmid1  Alexander Hapfelmeier3  Agnes S Meidert1  Julia S Prantner1  Julia Y Wagner2  | |
| [1] II. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, Ismaninger Strasse 22, 81675 München, Germany;Current affiliation: Department of Anesthesiology, Center of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Martinistrasse 52, 20246 Hamburg, Germany;Institut für Medizinische Statistik und Epidemiologie, Klinikum rechts der Isar der Technischen Universität München, Ismaninger Strasse 22, 81675 München, Germany | |
| 关键词: Volume clamp method; Photoplethysmography; Emergency medicine; Mean arterial pressure; Diastolic blood pressure; Systolic blood pressure; | |
| Others : 810382 DOI : 10.1186/1757-7241-22-8 |
|
| received in 2013-10-05, accepted in 2014-01-25, 发布年份 2014 | |
PDF
|
|
【 摘 要 】
Background
Monitoring cardiovascular function in acutely ill patients in the emergency department (ED) is of paramount importance. Arterial pressure (AP) is usually monitored using intermittent oscillometric measurements with an upper arm cuff. The vascular unloading technique (VUT) allows continuous noninvasive AP monitoring. In this study, we compare continuous AP measurements obtained by VUT with intermittent oscillometric AP measurements in ED patients. In addition, we aimed to investigate whether continuous noninvasive AP monitoring allows detection of relevant hypotensive episodes that might be missed with intermittent AP monitoring.
Methods
In a German university hospital, 130 ED patients who required AP monitoring were analyzed in this prospective method comparison study. Continuous AP monitoring was performed using VUT (CNAP technology; CNSystems Medizintechnik AG, Graz, Austria) over a 2-hour period. The oscillometric AP values were recorded simultaneously every 15 minutes for the comparison of both methods. For statistical evaluation, Bland-Altman plots accounting for repeated AP measurements per individual were used.
Results
The mean difference (±standard deviation) between AP measurements obtained by VUT and oscillometric AP measurements was -5 mmHg (±22 mmHg) for systolic AP (SAP), -2 mmHg (±15 mmHg) for diastolic AP (DAP), and -6 mmHg (±16 mmHg) for mean AP (MAP), respectively. In the interval between two oscillometric measurements, the VUT device detected hypotensive episodes (≥4 minutes) defined as either SAP <90 mmHg or MAP <65 mmHg in 30 patients and 16 patients, respectively. In 11 (SAP <90 mmHg) and 6 (MAP <65 mmHg) of these patients, hypotension was also detected by the subsequent intermittent oscillometric AP measurement.
Conclusions
VUT using the CNAP system for noninvasive continuous AP measurement shows reasonable agreement with intermittent oscillometric measurements in acutely ill ED patients. Continuous AP monitoring allows immediate recognition of clinically relevant hypotensive episodes, which are missed or only belatedly recognized with intermittent AP measurement.
【 授权许可】
2014 Wagner et al.; licensee BioMed Central Ltd.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 20140709041457584.pdf | 840KB | ||
| Figure 1. | 212KB | Image |
【 图 表 】
Figure 1.
【 参考文献 】
- [1]Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M: Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 2006, 34:1589-1596.
- [2]Frezza EE, Mezghebe H: Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients. Am Surg 1998, 64:127-131.
- [3]Fortin J, Marte W, Grullenberger R, Hacker A, Habenbacher W, Heller A, Wagner C, Wach P, Skrabal F: Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops. Comput Biol Med 2006, 36:941-957.
- [4]Biais M, Vidil L, Roullet S, Masson F, Quinart A, Revel P, Sztark F: Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery. Ann Fr Anesth Reanim 2010, 29:530-535.
- [5]Jeleazcov C, Krajinovic L, Munster T, Birkholz T, Fried R, Schuttler J, Fechner J: Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth 2010, 105:264-272.
- [6]Hahn R, Rinosl H, Neuner M, Kettner SC: Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™ 500) during general anaesthesia. Br J Anaesth 2012, 108:581-585.
- [7]Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R: Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth 2012, 108:202-210.
- [8]Jagadeesh AM, Singh NG, Mahankali S: A comparison of a continuous noninvasive arterial pressure (CNAP) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU. Ann Card Anaesth 2012, 15:180-184.
- [9]Eeftinck Schattenkerk DW, van Lieshout JJ, van den Meiracker AH, Wesseling KR, Blanc S, Wieling W, van Montfrans GA, Settels JJ, Wesseling KH, Westerhof BE: Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff. Am J Hypertens 2009, 22:378-383.
- [10]Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ: Noninvasive continuous arterial blood pressure monitoring with Nexfin(R). Anesthesiology 2012, 116:1092-1103.
- [11]Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C: Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth 2012, 109:609-615.
- [12]Janelle GM, Gravenstein N: An accuracy evaluation of the T-Line Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients. Anesth Analg 2006, 102:484-490.
- [13]Szmuk P, Pivalizza E, Warters RD, Ezri T, Gebhard R: An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia 2008, 63:307-312.
- [14]Dueck R, Goedje O, Clopton P: Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. J Clin Monit Comput 2012, 26:75-83.
- [15]Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W: The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med 2012, 38:1471-1477.
- [16]Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W: Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth 2013, 111:185-190.
- [17]Meidert AS, Huber W, Hapfelmeier A, Schofthaler M, Muller JN, Langwieser N, Wagner JY, Schmid RM, Saugel B: Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care 2013, 28:908-912.
- [18]Siebig S, Rockmann F, Sabel K, Zuber-Jerger I, Dierkes C, Brunnler T, Wrede CE: Continuous non-invasive arterial pressure technique improves patient monitoring during interventional endoscopy. Int J Med Sci 2009, 6:37-42.
- [19]Ilies C, Kiskalt H, Siedenhans D, Meybohm P, Steinfath M, Bein B, Hanss R: Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement. Br J Anaesth 2012, 109:413-419.
- [20]Penaz J, Voigt A, Teichmann W: Contribution to the continuous indirect blood pressure measurement. Z Gesamte Inn Med 1976, 31:1030-1033.
- [21]Bland JM, Altman DG: Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007, 17:571-582.
- [22]Nowak RM, Sen A, Garcia AJ, Wilkie H, Yang JJ, Nowak MR, Moyer ML: Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED. Am J Emerg Med 2011, 29:782-789.
- [23]ANSI/AAMI SP10:2002. American national standard for manual, electronic, or automated sphygmomanometers: Association for the Advancement of Medical Instrumentation. 2002.