期刊论文详细信息
Radiation Oncology
A technique for pediatric total skin electron irradiation
Timothy D Solberg2  Frederick H Hager2  Joseph P Dugas1  Brian A Hrycushko2  Qinan Bao2 
[1] Department of Physics and Astronomy, Louisiana State University, 202 Nicholson Hall, Baton Rouge, LA 70803, USA;Department of Radiation Oncology, University of Texas Southwestern Medical Center, 5801 Forest Park Rd, Dallas, TX 75390, USA
关键词: Leukemia cutis;    Quality assurance;    Commissioning;    Pediatric total skin electron irradiation;   
Others  :  1160876
DOI  :  10.1186/1748-717X-7-40
 received in 2011-12-19, accepted in 2012-03-20,  发布年份 2012
PDF
【 摘 要 】

Background

Total skin electron irradiation (TSEI) is a special radiotherapy technique which has generally been used for treating adult patients with mycosis fungoides. Recently, two infants presented with leukemia cutis isolated to the skin requiring TSEI. This work discusses the commissioning and quality assurance (QA) methods for implementing a modified Stanford technique using a rotating harness system to position sedated pediatric patients treated with electrons to the total skin.

Methods and Results

Commissioning of pediatric TSEI consisted of absolute calibration, measurement of dosimetric parameters, and subsequent verification in a pediatric patient sized cylindrical phantom using radiographic film and optically stimulated luminance (OSL) dosimeters. The depth of dose penetration under TSEI treatment condition was evaluated using radiographic film sandwiched in the phantom and demonstrated a 2 cm penetration depth with the maximum dose located at the phantom surface. Dosimetry measurements on the cylindrical phantom and in-vivo measurements from the patients suggested that, the factor relating the skin and calibration point doses (i.e., the B-factor) was larger for the pediatric TSEI treatments as compared to adult TSEI treatments. Custom made equipment, including a rotating plate and harness, was fabricated and added to a standard total body irradiation stand and tested to facilitate patient setup under sedated condition. A pediatric TSEI QA program, consisting of daily output, energy, flatness, and symmetry measurements as well as in-vivo dosimetry verification for the first cycle was developed. With a long interval between pediatric TSEI cases, absolute dosimetry was also repeated as part of the QA program. In-vivo dosimetry for the first two infants showed that a dose of ± 10% of the prescription dose can be achieved over the entire patient body.

Conclusion

Though pediatric leukemia cutis and the subsequent need for TSEI are rare, the ability to commission the technique on a modified TBI stand is appealing for clinical implementation and has been successfully used for the treatment of two pediatric patients at our institution.

【 授权许可】

   
2012 Bao et al; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150411082840782.pdf 327KB PDF download
Figure 3. 51KB Image download
Figure 2. 79KB Image download
Figure 1. 40KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

【 参考文献 】
  • [1]Jones G, Wilson LD, Fox-Goguen L: Total skin electron beam radiotherapy for patients who have mycosis fungoides. Hematol Oncol Clin North Am 2003, 17:1421-1434.
  • [2]Nisce LZ, Safai B, Poussin-Rosillo H: Once weekly total and subtotal skin electron beam therapy for Kaposi's sarcoma. Cancer 1981, 47:640-644.
  • [3]Leung SW, Hsu HC, Huang PH: Total skin electron beam irradiation in scleromyxoedema. Br J Radiol 1998, 71:84-86.
  • [4]Karzmark CJ: Large-field superficial electron therapy with linear accelerators. Br J Radiol 1964, 37:302-305.
  • [5]Wu JM, Yeh SA, Hsiao KY, Chao MM, Hargrove I: A conceptual design of rotating board technique for delivering total skin electron therapy. Med Phys 2010, 37:1449-1458.
  • [6]Podgorsak EB, Pla C, Pla M, Lefebvre PY, Heese R: Physical aspects of a rotational total skin electron irradiation. Med Phys 1983, 10:159-168.
  • [7]Earley L, Moeller J, O'Rear J, Leavitt DD: A method for total skin electron treatment for infants. Med Dosim 1995, 20:243-248.
  • [8]Wu JM, Leung SW, Wang CJ, Chui CS: Lying-on position of total skin electron therapy. Int J Radiat Oncol Biol Phys 1997, 39:521-528.
  • [9]Peters VG, Jaywant SM: Implementation of total skin electron therapy using an optional high dose rate mode on a conventional linear accelerator. Med Dosim 1995, 20:99-104.
  • [10]Trump JG, Wright KA, Evans WW, Anson JH, Hare HF, Fromer JL, Jacque G, Horne KW: High energy electrons for the treatment of extensive superficial malignant lesions. The American journal of roentgenology, radium therapy, and nuclear medicine 1953, 69:623-629.
  • [11]Karzmark CJ, Anderson J, Fessenden P, Svensson G, Buffa A, Khan FM, Wright KA: Total skin electron therapy: technique and dosimetry. Report of Task Group 30. 1987.
  • [12]Rubin CM, Arthur DC, Meyers G, McClain KL, Kim TH, Woods WG: Leukemia cutis treated with total skin irradiation. Cancer 1985, 55:2649-2652.
  • [13]Pepek JM, Paulino AC, Briones MA, Marcus RB Jr, Esiashvili N: Role of total skin electron beam therapy for leukemia cutis in pediatric patients. Pediatr Blood Cancer 2008, 50:1054-1055.
  • [14]Majdzadeh N, Jain SK, Murphy MC, Dugas JP, Hager F, Abdulrahman R: Total skin electron beam radiation in a pediatric patient with leukemia cutis: A case report. Journal of Pediatric Hematology/Oncology, in press.
  • [15]Dusenbery KE, Howells WB, Arthur DC, Alonzo T, Lee JW, Kobrinsky N, Barnard DR, Wells RJ, Buckley JD, Lange BJ, Woods WG: Extramedullary leukemia in children with newly diagnosed acute myeloid leukemia: a report from the children's cancer group. Journal of Pediatric Hematology/Oncology 2003, 25:760-768.
  • [16]Karzmark CJ, Loevinger R, Steele RE, Weissbluth M: A technique for large-field, superficial electron therapy. Radiology 1960, 74:633-644.
  • [17]Karzmark CJ: Physical aspects of whole-body superficial therapy with electrons. Front Radiat Ther Oncol 1968, 2:36-54.
  • [18]A protocol for the determination of absorbed dose from high-energy photon and electron beams. Task Group 21, Radiation Therapy Committee, AAPM Med Phys 1983, 10:741-771.
  • [19]Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, Rogers DW: AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999, 26:1847-1870.
  • [20]Van Der Merwe DG: Total skin electron therapy: A technique which can be implemented on a conventional electron linear accelerator. International Journal of Radiation Oncology Biology Physics 1993, 27:391-396.
  • [21]Niroomand-Rad A, Gillin MT, Komaki R, Kline RW, Grimm DF: Dose distribution in total skin electron beam irradiation using the six-field technique. Int J Radiat Oncol Biol Phys 1986, 12:415-419.
  • [22]Khan FM: Electron beam therapy. In The Physics of Radiation Therapy. 4th edition. Philadelphia, PA: Lippincott Williams & Wilkins; 2009:299-306.
  • [23]Esquivel C, Smith MS, Stathakis S, Gutierrez A, Shi C, Papanikolaou N: Total skin electron therapy skin dose validation using optically stimulated luminescent dosimeters [abstract]. AAPM Annual Meeting 2009. ID10746