Management of radiation oncology patients with a pacemaker or ICD: A new comprehensive practical guideline in The Netherlands
Lieselot van Erven5  Marleen MJ Ploegmakers4  Arnoud J van der Borden3  GJ Uiterwaal2  Ad JJ Maas2  Bing S Oei1  Joost L Knegjens1  Coen W Hurkmans1 
[1] Dutch Society of Radiotherapy and Oncology (NVRO), Utrecht, The Netherlands;Dutch Society of Specialists in Invasive Heart Stimulation (VITHaS), Utrecht, The Netherlands;Department of Radiation Oncology, University Medical Centre Groningen, Groningen, The Netherlands;Order of Medical Specialists (OMS), Utrecht, The Netherlands;Department of Cardiology, Leiden University Medical Centre, Leiden, The Netherlands
关键词: Risk management;    CIED;    Guideline;    Implantable cardioverter defibrillator;    Pacemaker;   
Others  :  1155020
DOI  :  10.1186/1748-717X-7-198
 received in 2012-07-24, accepted in 2012-11-21,  发布年份 2012
PDF
【 摘 要 】

Current clinical guidelines for the management of radiotherapy patients having either a pacemaker or implantable cardioverter defibrillator (both CIEDs: Cardiac Implantable Electronic Devices) do not cover modern radiotherapy techniques and do not take the patient’s perspective into account. Available data on the frequency and cause of CIED failure during radiation therapy are limited and do not converge. The Dutch Society of Radiotherapy and Oncology (NVRO) initiated a multidisciplinary task group consisting of clinical physicists, cardiologists, radiation oncologists, pacemaker and ICD technologists to develop evidence based consensus guidelines for the management of CIED patients. CIED patients receiving radiotherapy should be categorised based on the chance of device failure and the clinical consequences in case of failure. Although there is no clear cut-off point nor a clear linear relationship, in general, chances of device failure increase with increasing doses. Clinical consequences of device failures like loss of pacing, carry the most risks in pacing dependent patients. Cumulative dose and pacing dependency have been combined to categorise patients into low, medium and high risk groups. Patients receiving a dose of less than 2 Gy to their CIED are categorised as low risk, unless pacing dependent since then they are medium risk. Between 2 and 10 Gy, all patients are categorised as medium risk, while above 10 Gy every patient is categorised as high risk. Measures to secure patient safety are described for each category. This guideline for the management of CIED patients receiving radiotherapy takes into account modern radiotherapy techniques, CIED technology, the patients’ perspective and the practical aspects necessary for the safe management of these patients. The guideline is implemented in The Netherlands in 2012 and is expected to find clinical acceptance outside The Netherlands as well.

【 授权许可】

   
2012 Hurkmans et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150407111739604.pdf 299KB PDF download
Figure 1. 108KB Image download
【 图 表 】

Figure 1.

【 参考文献 】
  • [1]Crossley GH, Poole JE, Rozner MA, et al.: The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management this document was developed as a joint project with the American Society of Anesthesiologists (ASA), and in collaboration with the American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Heart Rhythm 2011, 8:1114-1154.
  • [2]Solan AN, Solan MJ, Bednarz G, et al.: Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy. Int J Radiat Oncol Biol Phys 2004, 59:897-904.
  • [3]Wadasadawala T, Pandey A, Agarwal JP, et al.: Radiation therapy with implanted cardiac pacemaker devices: a clinical and dosimetric analysis of patients and proposed precautions. Clin Oncol (R Coll Radiol) 2011, 23:79-85.
  • [4]Hudson F, Coulshed D, D'Souza E, et al.: Effect of radiation therapy on the latest generation of pacemakers and implantable cardioverter defibrillators: A systematic review. J Med Imaging Radiat Oncol 2010, 54:53-61.
  • [5]Snow JS, Kalenderian D, Colasacco JA, et al.: Implanted devices and electromagnetic interference: case presentations and review. J Invasive Cardiol 1995, 7:25-32.
  • [6]Last A: Radiotherapy in patients with cardiac pacemakers. Br J Radiol 1998, 71:4-10.
  • [7]Niehaus M, Tebbenjohanns J: Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators. Heart 2001, 86:246-248.
  • [8]Sundar S, Symonds RP, Deehan C: Radiotherapy to patients with artificial cardiac pacemakers. Cancer Treat Rev 2005, 31:474-486.
  • [9]Tondato F, Ng DW, Srivathsan K, et al.: Radiotherapy-induced pacemaker and implantable cardioverter defibrillator malfunction. Expert Rev Med Devices 2009, 6:243-249.
  • [10]Marbach JR, Sontag MR, Van DJ, et al.: Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine. Med Phys 1994, 21:85-90.
  • [11]Soejima T, Yoden E, NIshimura Y, et al.: Radiation therapy in patients with implanted cardiac pacemakers and implantable cardioverter defibrillators: a prospective survey in Japan. J Radiat Res 2011, 52:516-521.
  • [12]Lin JP, Chu TC, Lin SY, et al.: The measurement of photoneutrons in the vicinity of a Siemens Primus linear accelerator. Appl Radiat Isot 2001, 55:315-321.
  • [13]Nath R, Meigooni AS, King CR, et al.: Superheated drop detector for determination of neutron dose equivalent to patients undergoing high-energy x-ray and electron radiotherapy. Med Phys 1993, 20:781-787.
  • [14]Croshaw R, Kim Y, Lappinen E, et al.: Avoiding mastectomy: accelerated partial breast irradiation for breast cancer patients with pacemakers or defibrillators. Ann Surg Oncol 2011, 18:3500-3505.
  • [15]Kim Y, Arshoun Y, Trombetta MG: Pacemaker/implantable cardioverter-defibrillator dose in balloon high-dose-rate brachytherapy for breast cancer treatment. Brachytherapy 2012, 11:380-386.
  • [16]Diederich S, Lenzen H: Radiation exposure associated with imaging of the chest: comparison of different radiographic and computed tomography techniques. Cancer 2000, 89:2457-2460.
  • [17]Murphy MJ, Balter J, Balter S, et al.: The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. Med Phys 2007, 34:4041-4063.
  • [18]Rodriguez F, Filimonov A, Henning A, et al.: Radiation-induced effects in multiprogrammable pacemakers and implantable defibrillators. Pacing Clin Electrophysiol 1991, 14:2143-2153.
  • [19]John J, Kaye GC: Shock coil failure secondary to external irradiation in a patient with implantable cardioverter defibrillator. Pacing Clin Electrophysiol 2004, 27:690-691.
  • [20]Guidant Corporation Cardiac Rhythm Management Technical Services: The impact of therapeutic radiation on Guidant implantable pacemakers (IPMs) and implantable cardioverter defibrillators (ICDs). 2003, 1-6. [Revision L02/13/03]
  • [21]Boston Scientific: Therapeutic Radiation and Implantable Pacemakers and Defibrillators. 2008.
  • [22]Medtronic USA Inc.: Standard Letter, THERAPEUTIC RADIATION. [Rev. E 22-12-0008]
  • [23]St.jude Medical: Effects of Therapeutic Radiation on St. Jude Medical Implantable Cardiac Rhythm Devices. 2008. [Technical Services Rev 11/08]
  • [24]Uiterwaal H, Springorum BG, Scheepers E, et al.: Interference detection in implantable defibrillators induced by therapeutic radiation therapy. Neth Heart J 2006, 14:330-334.
  • [25]Hurkmans CW, Scheepers E, Springorum BG, et al.: Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators. Int J Radiat Oncol Biol Phys 2005, 63:282-289.
  • [26]Hurkmans CW, Scheepers E, Springorum BG, et al.: Influence of radiotherapy on the latest generation of pacemakers. Radiother Oncol 2005, 76:93-98.
  • [27]Zweng A, Schuster R, Hawlicek R, et al.: Life-threatening pacemaker dysfunction associated with therapeutic radiation: a case report. Angiology 2009, 60:509-512.
  • [28]Mouton J, Haug R, Bridier A, et al.: Influence of high-energy photon beam irradiation on pacemaker operation. Phys Med Biol 2002, 47:2879-2893.
  • [29]McCollough CH, Zhang J, Primak AN, et al.: Effects of CT irradiation on implantable cardiac rhythm management devices. Radiology 2007, 243:766-774.
  • [30]Souliman SK, Christie J: Pacemaker failure induced by radiotherapy. Pacing Clin Electrophysiol 1994, 17:270-273.
  • [31]Kapa S, Fong L, Blackwell CR, et al.: Effects of scatter radiation on ICD and CRT function. Pacing Clin Electrophysiol 2008, 31:727-732.
  • [32]Hashimoto T, Isobe T, Hashii H, et al.: Influence of secondary neutrons induced by proton radiotherapy for cancer patients with implantable cardioverter defibrillators. Radiat Oncol 2012, 7:10. 10 BioMed Central Full Text
  • [33]Franco L, Gomez F, Iglesias A, Pardo J, Pazos A, Pena J, Zapata M: SEUs on commercial SRAM induced by low energy neutrons produced at a clinical linac facility. 2005. [Radecs]
  • [34]Wilkinson JD, Bounds C, Brown T, Gerbi BJ, Peltier J: Cancer-Radiotherapy Equipment as a Cause of Soft Errors in Electronic Equipment. Device and Materials Reliability. 2005, 449-451. [IEEE Transactions]
  • [35]Hashii H, Hashimoto T, Okawa A, et al.: Comparison of the Effects of High-Energy Photon Beam Irradiation (10 and 18 MV) on 2 Types of Implantable Cardioverter-Defibrillators. Int J Radiat Oncol Biol Phys 2012, 78:S71-S72.
  • [36]Ferrara T, Baiotto B, Malinverni G, et al.: Irradiation of pacemakers and cardio-defibrillators in patients submitted to radiotherapy: a clinical experience. Tumori 2010, 96:76-83.
  • [37]Gelblum DY, Amols H: Implanted cardiac defibrillator care in radiation oncology patient population. Int J Radiat Oncol Biol Phys 2009, 73:1525-1531.
  • [38]Lau DH, Wilson L, Stiles MK, et al.: Defibrillator reset by radiotherapy. Int J Cardiol 2008, 130:e37-e38.
  • [39]Nemec J: Runaway implantable defibrillator–a rare complication of radiation therapy. Pacing Clin Electrophysiol 2007, 30:716-718.
  • [40]Tsekos A, Momm F, Brunner M, et al.: The cardiac pacemaker patient–might the pacer be directly irradiated? Acta Oncol 2000, 39:881-883.
  • [41]Hauser RG, Hayes DL, Epstein AE, et al.: Multicenter experience with failed and recalled implantable cardioverter-defibrillator pulse generators. Heart Rhythm 2006, 3:640-644.
  • [42]Lelakowski J, Majewski J, Bednarek J, et al.: Pacemaker dependency after pacemaker implantation. Cardiol J 2007, 14:83-86.
  • [43]Korantzopoulos P, Letsas KP, Grekas G, et al.: Pacemaker dependency after implantation of electrophysiological devices. Europace 2009, 11:1151-1155.
  • [44]Vardas PE, Auricchio A, Blanc JJ, et al.: Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in collaboration with the European Heart Rhythm Association. Eur Heart J 2007, 28:2256-2295.
  • [45]Sears SF, Hauf JD, Kirian K, et al.: Posttraumatic stress and the implantable cardioverter-defibrillator patient: what the electrophysiologist needs to know. Circ Arrhythm Electrophysiol 2011, 4:242-250.
  • [46]Habibovic M, van den Broek KC, Alings M, et al.: Posttraumatic stress 18 months following cardioverter defibrillator implantation: shocks, anxiety, and personality. Health Psychol 2012, 31:186-193.
  • [47]Bostwick JM, Sola CL: An updated review of implantable cardioverter/defibrillators, induced anxiety, and quality of life. Heart Fail Clin 2011, 7:101-108.
  • [48]Wilkoff BL, Love CJ, Byrd CL, et al.: Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm 2009, 6:1085-1104.
  • [49]de Bie MK, van Rees JB, Thijssen J, et al.: Cardiac device infections are associated with a significant mortality risk. Heart Rhythm 2012, 9:494-498.
  • [50]Gould PA, Krahn AD: Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories. JAMA 2006, 295:1907-1911.
  • [51]Borleffs CJ, Thijssen J, de Bie MK, et al.: Recurrent implantable cardioverter-defibrillator replacement is associated with an increasing risk of pocket-related complications. Pacing Clin Electrophysiol 2010, 33:1013-1019.
  • [52]Munshi A, Wadasadawala T, Sharma PK, et al.: Radiation therapy planning of a breast cancer patient with in situ pacemaker–challenges and lessons. Acta Oncol 2008, 47:255-260.
  • [53]Wilm M, Kronholz HL, Koch T: Beeinflussung programmierbarer schrittmacher durch therapeutische bestralungen. Strahlenther Onkol 1994, 170:225-231.
  文献评价指标  
  下载次数:7次 浏览次数:18次