Radiation Oncology
Stereotactic body radiation therapy for liver tumours using flattening filter free beam: dosimetric and technical considerations
Marta Scorsetti1  Luca Cozzi2  Piera Navarria1  Antonella Fogliata2  Francesca Lobefalo1  Angelo Tozzi1  Stefano Arcangeli1  Chiara Pellegrini1  Filippo Alongi1  Maddalena Catalano1  Giacomo Reggiori1  Simona Castiglioni1  Pietro Mancosu1 
[1] IRCCS Istituto Clinico Humanitas, Rozzano (Milano), Italy;Oncology Institute of Southern Switzerland, Bellinzona, Switzerland
关键词: TrueBeam;    RapidArc;    flattening filter free;    liver;    SBRT;   
Others  :  1160900
DOI  :  10.1186/1748-717X-7-16
 received in 2011-12-30, accepted in 2012-02-01,  发布年份 2012
PDF
【 摘 要 】

Purpose

To report the initial institute experience in terms of dosimetric and technical aspects in stereotactic body radiation therapy (SBRT) delivered using flattening filter free (FFF) beam in patients with liver lesions.

Methods and Materials

From October 2010 to September 2011, 55 consecutive patients with 73 primary or metastatic hepatic lesions were treated with SBRT on TrueBeam using FFF beam and RapidArc technique. Clinical target volume (CTV) was defined on multi-phase CT scans, PET/CT, MRI, and 4D-CT. Dose prescription was 75 Gy in 3 fractions to planning target volume (PTV). Constraints for organs at risk were: 700 cc of liver free from the 15 Gy isodose, Dmax < 21 Gy for stomach and duodenum, Dmax < 30 Gy for heart, D0.1 cc < 18 Gy for spinal cord, V15 Gy < 35% for kidneys. The dose was downscaled in cases of not full achievement of dose constraints. Daily cone beam CT (CBCT) was performed.

Results

Forty-three patients with a single lesion, nine with two lesions and three with three lesions were treated with this protocol. Target and organs at risk objectives were met for all patients. Mean delivery time was 2.8 ± 1.0 min. Pre-treatment plan verification resulted in a Gamma Agreement Index of 98.6 ± 0.8%. Mean on-line co-registration shift of the daily CBCT to the simulation CT were: -0.08, 0.05 and -0.02 cm with standard deviations of 0.33, 0.39 and 0.55 cm in, vertical, longitudinal and lateral directions respectively.

Conclusions

SBRT for liver targets delivered by means of FFF resulted to be feasible with short beam on time.

【 授权许可】

   
2012 Mancosu et al; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150411083532183.pdf 2610KB PDF download
Figure 3. 54KB Image download
Figure 2. 186KB Image download
Figure 1. 80KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

【 参考文献 】
  • [1]Timmerman RD, Kavanagh BD, Cho LC, Papiez L, Xing L: Stereotactic body radiation therapy in multiple organ sites. J Clin Oncol 2007, 25:947-952.
  • [2]Macdermed DM, Weichselbaum RR, Salama JK: A rationale for the targeted treatment of oligometastases with radiotherapy. J Surg Oncol 2008, 98:202-206.
  • [3]Schefter TE, Kavanagh BD, Timmerman RD, Cardenes HR, Baron A, Gaspar LE: A phase I trial of stereotactic body radiation therapy (SBRT) for liver metastases. Int J Radiation Oncol Biol Phys 2005, 62:1371-1378.
  • [4]Katz AW, Carey-Sampson M, Muhs AG, Milano MT, Schell MC, Okunieff P: Hypofractionated stereotactic body radiation therapy (SBRT) for limited hepatic metastases. Int J Radiation Oncology Biol Phys 2007, 67:793-798.
  • [5]Scorsetti M, Fogliata A, Castiglioni S, et al.: Early clinical experience with volumetric modulated arc therapy in head and neck cancer patients. Radiation Oncology 2010, 5:93. BioMed Central Full Text
  • [6]Scorsetti M, Navarria P, Alongi F, et al.: Large volume unresectable locally advanced non-small cell lung cancer: acute toxicity and initial outcome results with rapid arc. Radiation Oncology 2010, 5:94. BioMed Central Full Text
  • [7]Scorsetti M, Bignardi M, Clivio A, et al.: Volumetric modulation arc radiotherapy compared with static gantry intensity-modulated radiotherapy for malignant pleural mesothelioma tumor: a feasibility study. Int J Radiat Oncol Biol Phys 2010, 77:942-949.
  • [8]Fogliata A, Bergstrom S, Cafaro I, et al.: Cranio-spinal irradiation with volumetric modulated arc therapy: a multi-institutional treatment experience. Radiotherapy and Oncology 2011, 99:79-85.
  • [9]Mancosu P, Cozzi L, Fogliata A, et al.: Collimator angle influence on dose distribution optimization for vertebral metastases using volumetric modulated arc therapy. Med Phys 2010, 37:4133-4137.
  • [10]Mancosu P, Navarria P, Bignardi M, et al.: Re-irradiation of metastatic spinal cord compression: a feasibility study by volumetric-modulated arc radiotherapy for in-field recurrence creating a dosimetric hole on the central canal. Radiotherapy and Oncology 2010, 94:67-70.
  • [11]Fogliata A, Cozzi L, Clivio A, et al.: Preclinical assessment of volumetric modulated arc therapy for total marrow irradiation. Int J Radiat Oncol Biol Phys 2011, 80:628-636.
  • [12]Mancosu P, Navarria P, Castagna L, et al.: Anatomy driven optimisation strategy for total marrow irradiation with a volumetric modulated arc therapy technique. JACMP 2011, in press.
  • [13]Scorsetti M, Mancosu P, Navarria P, et al.: Stereotactic body radiation therapy (SBRT) for adrenal metastases: a feasibility study of advanced techniques with modulated photons and protons. Strahlenther Onkol 2011, 187:238-244.
  • [14]Bignardi M, Cozzi L, Fogliata A, Lattuada P, Mancosu P, Navarria P, et al.: Critical appraisal of volumetric modulated arc therapy in stereotactic body radiation therapy for metastases to abdominal lymph nodes. Int J Radiat Oncol Biol Phys 2009, 75:1570-1577.
  • [15]Bignardi M, Navarria P, Mancosu P, Cozzi L, Fogliata A, Tozzi A, et al.: Clinical outcome of hypofractionated stereotactic radiotherapy for abdominal lymph node metastases. Int J Radiati Oncol Biol Phys Vol 2010, in press.
  • [16]Scorsetti M, Bignardi M, ALongi F, Fogliata A, Mancosu P, Navarria P, Castiglioni S, Pentimalli S, Tozzi A, Cozzi L: Stereotactic body radiation therapy for abdominal targets using volumetric intensity modulated arc therapy with RapidArc: feasibility and clinical preliminary results. Acta Oncol 2011, 50(4):528-38.
  • [17]Hrbacek J, Lang S, Klöck S: Comissioning of photon beams of a flattening filter-free linear accelerator and the accuracy of beam modelling using an anisotropic analytical algorithm. Int J Radiat Oncol Biol Phys, in press.
  • [18]Vassiliev ON, Titt U, Pönisch F, Kry SF, Mohan R, Gillin MT: Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol 2006, 51:1907-17.
  • [19]Kry SF, Howell RM, Titt U, Salehpour M, Mohan R, Vassiliev ON: Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. Med Phys 2008, 35(5):1906-11.
  • [20]Pönisch F, Titt U, Vassiliev ON, Kry SF, Mohan R: Properties of unflattened photon beams shaped by a multileaf collimator. Med Phys 2006, 33(6):1738-46.
  • [21]Vassiliev ON, Titt U, Kry SF, Pönisch F, Gillin MT, Mohan R: Monte Carlo study of photon fields from a flattening filter-free clinical accelerator. Med Phys 2006, 33(4):820-7.
  • [22]Georg D, Knöös T, McClean B: Current status and future perspective of flattening filter free photon beams. Med Phys 2011, 38(3):1280-93.
  • [23]Kragl G, Baier F, Lutz S, Albrich D, Dalaryd M, Kroupa B, Wiezorek T, Knöös T, Georg D: Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses. Z Med Phys 2010, in press.
  • [24]Vassiliev ON, Kry SF, Chang JY, Balter PA, Titt U, Mohan R: Stereotactic radiotherapy for lung cancer using a flattening filter free Clinac. J Appl Clin Med Phys 2009, 10(1):2880..
  • [25]Scorsetti M, Alongi F, Castiglioni S, Clivio A, Fogliata A, Lobefalo F, Mancosu P, Navarria P, Palumbo V, Pellegrini C, Pentimalli S, Reggiori G, Ascolese AM, Roggio A, Arcangeli S, Tozzi A, Vanetti E, Cozzi L: Feasibility and early clinical assessment of flattening filter free (FFF) based stereotactic body radiotherapy (SBRT) treatments. Radiat Oncol 2011, 6:113. BioMed Central Full Text
  • [26]Eccles CL, Dawson L, Moseley JL, Brock KK: Intrafraction liver shape variability and impact on GTV position during liver stereotactic radiotherapy using abdominal compression. Int J Radiat Oncol Biol Phys 2011, 80(3):938-946.
  • [27]Reggiori G, Mancosu P, Tozzi A, et al.: Cone beam CT pre- and post-daily treatment for assessing geometrical and dosimetric intrafraction variability during radiotherapy of prostate cancer. J Appl Clin Med Phys 2010, 12:3371.
  • [28]RAS-trial radiofrequency ablation versus stereotactic body radiation therapy for colorectal liver metastases: a randomized trial Available at: http://www.livertumor.dk/RAS-trial.html webcite
  • [29]Milano MT, Constine LS, Okunieff P: Normal tissue toxicity after small field hypofractionated stereotactic body radiation. Radiation Oncology 2008, 3:36. BioMed Central Full Text
  • [30]Van Herk M, Remeijer P, Rasch C, Lebesque JV: The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys 2000, 47(4):1121-35.
  • [31]Low DA, Harms WB, Mutic S, Purdy JA: A technique for quantitative evaluation of dose distributions. Med Phys 1998, 25:656-661.
  • [32]Reggiori G, Mancosu P, Castiglioni S, et al.: Can volumetric modulated arc therapy with flattening filter free beams play a role in stereotactic body radiotherapy for liver lesions? Med Phys 2012, in press.
  文献评价指标  
  下载次数:28次 浏览次数:7次