| Radiation Oncology | |
| Influence of secondary neutrons induced by proton radiotherapy for cancer patients with implantable cardioverter defibrillators | |
| Hideyuki Sakurai2  Kazutaka Aonuma1  Takeji Sakae2  Koji Tsuboi2  Toshiyuki Okumura2  Hiroshi Tada1  Hiroaki Kumada2  Haruko Hashii2  Tomonori Isobe2  Takayuki Hashimoto2  | |
| [1] Department of Cardiovascular Medicine, Graduate School of Comprehensive Human Sciences, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575, Japan;Department of Radiation Oncology, Graduate School of Comprehensive Human Sciences, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575, Japan | |
| 关键词: Monte-Carlo simulation; Soft error; Implantable cardioverter defibrillator; Secondary neutrons; Proton radiotherapy; | |
| Others : 1160906 DOI : 10.1186/1748-717X-7-10 |
|
| received in 2011-08-17, accepted in 2012-01-29, 发布年份 2012 | |
PDF
|
|
【 摘 要 】
Background
Although proton radiotherapy is a promising new approach for cancer patients, functional interference is a concern for patients with implantable cardioverter defibrillators (ICDs). The purpose of this study was to clarify the influence of secondary neutrons induced by proton radiotherapy on ICDs.
Methods
The experimental set-up simulated proton radiotherapy for a patient with an ICD. Four new ICDs were placed 0.3 cm laterally and 3 cm distally outside the radiation field in order to evaluate the influence of secondary neutrons. The cumulative in-field radiation dose was 107 Gy over 10 sessions of irradiation with a dose rate of 2 Gy/min and a field size of 10 × 10 cm2. After each radiation fraction, interference with the ICD by the therapy was analyzed by an ICD programmer. The dose distributions of secondary neutrons were estimated by Monte-Carlo simulation.
Results
The frequency of the power-on reset, the most serious soft error where the programmed pacing mode changes temporarily to a safety back-up mode, was 1 per approximately 50 Gy. The total number of soft errors logged in all devices was 29, which was a rate of 1 soft error per approximately 15 Gy. No permanent device malfunctions were detected. The calculated dose of secondary neutrons per 1 Gy proton dose in the phantom was approximately 1.3-8.9 mSv/Gy.
Conclusions
With the present experimental settings, the probability was approximately 1 power-on reset per 50 Gy, which was below the dose level (60-80 Gy) generally used in proton radiotherapy. Further quantitative analysis in various settings is needed to establish guidelines regarding proton radiotherapy for cancer patients with ICDs.
【 授权许可】
2012 Hashimoto et al; licensee BioMed Central Ltd.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 20150411083644342.pdf | 1680KB | ||
| Figure 4. | 116KB | Image | |
| Figure 3. | 109KB | Image | |
| Figure 2. | 68KB | Image | |
| Figure 1. | 50KB | Image |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
【 参考文献 】
- [1]Teshima T, Numasaki H, Shibuya H, Nishio M, Ikeda H, Ito H, Sekiguchi K, Kamikonya N, Koizumi M, Tago M, Nagata Y, Masaki H, Nishimura T, Yamada S: Japanese structure survey of radiation oncology in 2005 based on institutional stratification of patterns of care study. Int J Radiat Oncol Biol Phys 2008, 72:144-152.
- [2]Solan AN, Solan MJ, Bednarz G, Goodkin MB: Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy. Int J Radiat Oncol Biol Phys 2004, 59:897-904.
- [3]Marbach JR, Sontag MR, Van Dyk J, Wolbarst AB: Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine. Med Phys 1994, 21:85-90.
- [4]Kapa S, Fong L, Blackwell CR, Herman MG, Schomberg PJ, Hayes DL: Effects of scatter radiation on ICD and CRT function. Pacing Clin Electrophysiol 2008, 31:727-732.
- [5]Hurkmans CW, Scheepers E, Springorum BG, Uiterwaal H: Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators. Int J Radiat Oncol Biol Phys 2005, 63:282-289.
- [6]Yan X, Titt U, Koehler AM, Newhauser WD: Measurement of neutron dose equivalent to proton therapy patients outside of the proton radiation field. Nucl Instrum Methods Phys Res Sect A-Accel Spectrom Dect Assoc Equip 2002, 476:429-434.
- [7]Morávek Z, Bogner L: Analysis of the physical interactions of therapeutic proton beams in water with the use of Geant4 Monte Carlo calculations. Z Med Phys 2009, 19:174-181.
- [8]Pérez-Andújar A, Newhauser WD, DeLuca PM Jr: Neutron production from beam-modifying devices in a modern double scattering proton therapy beam delivery system. Phys Med Biol 2009, 54:993-1008.
- [9]Yonai S, Matsufuji N, Kanai T, Matsui Y, Matsushita K, Yamashita H, Numano M, Sakae T, Terunuma T, Nishio T, Kohno R, Akagi T: Measurement of neutron ambient dose equivalent in passive carbon-ion and proton radiotherapies. Med Phys 2008, 35:4782-4792.
- [10]Dowdell S, Clasie B, Wroe A, Guatelli S, Metcalfe P, Schulte R, Rosenfeld A: Tissue equivalency of phantom materials for neutron dosimetry in proton therapy. Med Phys 2009, 36:5412-5419.
- [11]Oshiro Y, Sugahara S, Noma M, Sato M, Sakakibara Y, Sakae T, Hayashi Y, Nakayama H, Tsuboi K, Fukumitsu N, Kanemoto A, Hashimoto T, Tokuuye K: Proton beam therapy interference with implanted cardiac pacemakers. Int J Radiat Oncol Biol Phys 2008, 72:723-727.
- [12]Blake SW: Advances in treatment planning dosimetry for a high energy neutron beam. Br J Radiol 1991, 64:341-349.
- [13]Isobe T, Kumada H, Takada K, Hashimoto T, Hashii H, Shida K, Tadano K, Fujibuchi T, Hanmura M, Sakurai H, Sakae T: Effects of secondary neutron beam generated in radiotherapy on electronic medical devices. Prog Nucl Sci Technol 2011, 2:524-529.
- [14]Kumada H, Sakae T, Saito K, Isobe T, Hashimoto T, Sakurai H: Application of the multi-model Monte-Carlo treatment planning system combined with PHITS to proton radiotherapy. Prog Nucl Sci Technol 2011, 2:213-218.
- [15]Hurkmans CW, Scheepers E, Springorum BG, Uiterwaal H: Influence of radiotherapy on the latest generation of pacemakers. Radiother Oncol 2005, 76:93-98.
- [16]Last A: Radiotherapy in patients with cardiac pacemakers. Br J Radiol 1998, 71:4-10.
- [17]Marbach JR: Recommendation precautions in the management of radiation oncology patients with implanted cardiac pacemaker. ASTRO Newsletter 1990, VII:7-8.
- [18]Lau DH, Wilson L, Stiles MK, John B, Shashidhar , Dimitri H, Brooks AG, Young GD, Sanders P: Defibrillator reset by radiotherapy. Int J Cardiol 2008, 130:e37-38.
- [19]Raitt MH, Stelzer KJ, Laramore GE, Bardy GH, Dolack GL, Poole JE, Kudenchuk PJ: Runaway pacemaker during high-energy neutron radiation therapy. Chest 1994, 106:955-957.
- [20]Tayama R, Fujita Y, Tadokoro M, Fujimaki H, Sakae T, Terunuma T: Measurement of neutron dose distribution for a passive scattering nozzle at the Proton Medical Research Center (PMRC). Nucl Instrum Methods Phys Res Sect A-Accel Spectrom Dect Assoc Equip 2006, 564:532-536.
- [21]Schneider U, Agosteo S, Pedroni E, Besserer J: Secondary neutron dose during proton therapy using spot scanning. Int J Radiat Oncol Biol Phys 2002, 53:244-251.
- [22]Rodriguez F, Filimonov A, Henning A, Coughlin C, Greenberg M: Radiation-induced effects in multiprogrammable pacemakers and implantable defibrillators. Pacing Clin Electrophysiol 1991, 14:2143-2153.
PDF